International
Tables for Crystallography Volume C Mathematical, physical and chemical tables Edited by E. Prince © International Union of Crystallography 2006 |
International Tables for Crystallography (2006). Vol. C. ch. 7.1, pp. 625-626
|
The amplitude of the signal due to a single photon is usually a function of the photon energy. The variance in this amplitude, or the full width at half-maximum (FWHM) of the pulse-height spectrum, for a monoenergetic input, depends on the statistics of the detection process. A sharp pulse-height-distribution (PHD) curve may permit simultaneous multi-wavelength measurements with a suitable counter, or at least afford a reduction of the background by pulse-height discrimination. In an analogue detector, the variance in the primary amplitude affects the DQE (Arndt & Gilmore, 1979).
References
Arndt, U. W. & Gilmore, D. J. (1979). X-ray television area detectors for macromolecular structural studes with synchrotron radiation sources. J. Appl. Cryst. 12, 1–9.Google Scholar