Section 7.1.7. X-ray-sensitive TV cameras
J. Chikawac
|
High-resolution X-ray imaging systems are required for the topographic study of spatial change in crystal structures, such as that which takes place in phase transformations. From this point of view, high-resolution TV camera tubes will be described.
Video displays of X-ray and optical images have different features. Although X-ray photon energies are very large, the intensities available in X-ray diffraction are extremely low compared with optical images. Therefore, the photon noise resulting from the statistical fluctuation of the number of photons incident upon an image system gives a detection limit of the image.
Consider the case of defects in a crystal viewed by an imaging system: νp photons s−1 mm−2 are diffracted from the perfect region of a crystal, and are absorbed by the X-ray-sensing layer of the system. An absorbed photon produces a mean number of electrons or visible photons, each of which may be rescattered to produce a mean number of electrons or photons. By repeating s such processes, the mean signal height is obtained from each square-shaped picture element δ × δ mm for t s. The value of δ may be taken as the limiting resolution of the system. Since owing to the large photon energy and the values of , are considered to be less than 100, the photon noise σp as a standard deviation of Sp is given by (Arcese, 1964) In order to recognize the defect image in the perfect region, the signal-to-noise ratio (SNR) of the image must be where Sd = signal height of the defect image, C = contrast of the defect image, ; and κ = threshold SNR of 1–5 (Rose, 1948).
Sensitivity is expressed by the intensity required for a certain signal height to the noise height of the imaging system and is proportional to the value of . As is seen from (7.1.7.3), however, the ratio of the signal to the photon noise depends only on the absorption efficiency q. For example, a 100 µm thick silicon-diode array (Rozgonyi, Haszlo & Statile, 1970) and 15 µm thick lead oxide camera tubes (Chikawa, 1974) have similar sensitivities for Mo Kα radiation, but q = 0.15 and q = 0.6, respectively; the former obtains the sensitivity with a higher conversion efficiency of absorbed photon energy and gives a (0.15/0.6)1/2 = 0.5 times lower SNR.
It is clear from the foregoing discussion that imaging systems should be evaluated by three factors: resolution, integration time for observation, and SNR.
There are two kinds of imaging system (e.g. Green, 1977); one is a direct method in which X-ray input images are converted directly into video signals and the other is an indirect method in which X-ray images are first converted into visible-light images to be observed by the usual electro-optical system. In the latter, phosphor screens are widely used and the visible images can be magnified by a lens. However, the resolution is limited by the thickness of the phosphor screen that is required for a satisfactory absorption efficiency. In contrast, the direct method provides a high resolution.
Various types of indirect method are discussed in Subsections 2.7.5.2 and 7.1.6.5. Here only the direct method will be described in some detail.
A direct method using X-ray TV camera tubes has been used for real-time topography (Chikawa & Matsui, 1994). The construction and operation of an X-ray TV camera tube are similar to those of the conventional video pick-up tube, except for a beryllium window, as shown schematically in Fig. 7.1.7.1. The popular one is a vacuum glass tube which is 15 cm long and 25 mm in diameter. An X-ray sensing photoconductive layer such as Se (McMaster, Photen & Mitchell, 1967), or PbO (Chikawa, 1974) is placed on the inner surface of the window at one end of the tube and is scanned by a narrow electron beam from a cathode on the opposite end to convert image charges on the photoconductive layer into video signals. Therefore, the resolution depends upon the characteristics of the photoconductive layer and the electron-beam size on the layer.
The camera tube with a PbO photoconductive layer has absorption efficiencies of q = 0.6 for Mo Kα and 0.8 for Cu Kα, using a PbO layer with a resolution-limited maximum thickness of 15 µm (Chikawa, 1974).
Dislocations in a silicon crystal were observed with the PbO camera tube and a high-power X-ray generator (tube voltage: 60 kV, current: 0.5 A) under the conditions νp = 4 × 1011 Mo Kα1 photons s−1 m−2, C = 0.5 and δ = 30 µm (Chikawa, 1974) in (7.1.7.3). In order to improve the resolution further, q must be as high as possible even when is increased by 10–100 by use of synchrotron radiation, because decreases with improving resolution (decreasing δ). (Note that C is not expected to increase much with improving resolution.) For example, to keep at the same level as the previous case, q should be 0.7, say, for νp = 1013 photons s−1 m−2 and δ = 6 µm. However, resolution and detection efficiency are, in general, mutually exclusive; the resolution of X-ray sensing photoconductive or phosphor materials is determined by the sizes of their grains and their sensitivities decrease with decreasing grain sizes. High resolution, without sacrificing detection efficiency, has been obtained with an amorphous Se–As alloy photoconductive layer, which has the advantage that no degradation of resolution occurs on increasing the layer thickness (Chikawa, Sato, Kawamura, Kuriyama, Yamashita & Goto, 1986). A layer with a thickness of 20 µm has an absorption efficiency of q = 0.52 for Mo Kα. The resolution of these tubes is shown with their modulation transfer functions in Fig. 7.1.7.2. The limiting resolution of 6 µm was obtained by making the scanning electron beam narrower with a diode-type electron gun have a barium-impregnated tungsten cathode (DIS type) (Chikawa, 1999). The Se–As layer shows very low lag characteristics (less than 1% after one frame).
This type of camera tube was developed primarily for TV broadcasting use and named `Saticon' as the acronym for the components of the photoconductor, Se, As and Te. (Te enhances the sensitivity to red light.)
Such camera tubes have excellent linearity. The output signal current is proportional to the incident X-ray intensity up to 4 µA. The amplifier is always improving in signal-to-noise ratio, and, at present, the noise level of video amplifiers is ∼0.2 nA; it is equivalent to an input intensity of ∼1011 Mo Kα photons m−2 s−1 in the US/Japan standard scanning system (30 frames s−1).
Therefore, it is important to increase conversion efficiency . With increasing voltage applied on the photoconductive layer, the signal current is saturated (in Fig. 7.1.7.1, all the holes produced by an incident photon are collected), and then increases again further by avalanche amplification, as shown in Fig. 7.1.7.3; holes accelerated by a strong electric field cause their multiplication (Sato, Maruyama, Goto, Fujimoto, Shidara, Kawamura, Hirai, Sakai & Chikawa, 1993). It was referred to as `HARP' (high-gain avalanche-rushing amorphous photoconductor). Together with the signal current, the dark current also increases as shown in Fig. 7.1.7.3. By allowing it to increase to the noise level of the video amplifier, an order of magnitude higher SNR can be achieved. Consequently, individual X-ray photons can be imaged as spots with a size resulting from the point-spread function, unless they are absorbed near the back surface of the photoconductive layer. For X-ray detection, a thick HARP layer should be employed with a very high applied voltage, and stable operation with avalanche amplification was confirmed for a 25 µm thick HARP layer. These pilot tubes were fabricated with a conventional electron gun and had a resolution of about 25 µm.
In general, avalanche amplification results in degradation of spatial resolution and has been used for zero-dimensional detection such as solid-state detectors. To make two-dimensional detection, isolation of each picture element is required. For the HARP, however, no appreciable degradation of resolution due to avalanche amplification was confirmed with a DIS-type tube having an 8 µm thick HARP layer by using visible light through a glass window. X-ray sensitive DIS-type tubes are now commercially available.
The resolution δ and integration time t should be selected appropriately according to experimental requirements (Chikawa, 1980). For example, when topographic images of a single dislocation in silicon were observed with t = 1/30 s by synchrotron radiation, their contrast and SNR were found to be C = 0.5 and RI = 20 for δ = 30 µm, and C = 1 and RI = 8 for δ = 6 µm. Since the SNR is desired to be 100, the integration time should be as large as possible unless images of moving objects are degraded. Digital image processing (Heynes, 1977) enables one to adjust the integration time easily. As an example, a noise reducer (McMann, Kreinik, Moore, Kaiser & Rossi, 1978; Rossi, 1978) is shown in Fig. 7.1.7.4. The video signal is sampled and digitized by the A/D converter and the digital video is sent to the adder and thence to the memory. Image information in the memory is continually sent both to the adder through the multiplier for combination with incoming data and to the display through the D/A converter. The weighting of new to old data is made by changing the factor k of the multiplier in the range . For k = 0, the original input image is displayed. In the range , a sliding summation of successive frames is displayed, and the SNR is improved by a factor of . The factor k can be adjusted automatically by detecting the difference between successive frames.
Using a HARP tube, the SNR can be improved without integration of the amplifier noise by image processing, and topographs were displayed with an intensity of photons s−1 m−2 by a conventional X-ray generator.
Acquisition of extremely low intensity images, dramatic improvements in SNR via frame integration, and isolation and enhancement of selected-contrast ranges are possible by digital image processing (Chikawa & Kuriyama, 1991).
References
Arcese, A. (1964). A note on Poisson branching processes with reference to photon–electron converters. Appl. Opt. 3, 435.Google ScholarChikawa, J. (1974). Technique for the video display of X-ray topographic images and its application to the study of crystal growth. J. Cryst. Growth, 24/25, 61–68.Google Scholar
Chikawa, J. (1980). Laboratory techniques for transmission X-ray topography. Characterization of crystal growth defects by X-ray methods, edited by B. K. Tanner & D. K. Bowen, Chap. 15, pp. 368–400. New York: Plenum.Google Scholar
Chikawa, J. (1999). Live X-ray topography and crystal growth of silicon. Jpn. J. Appl. Phys. 3, 4619–4631.Google Scholar
Chikawa, J. & Kuriyama, M. (1991). Topography. Handbook on synchrotron radiation, Vol. 3, edited by G. Brown & D. E. Moncton, Chap. 10, pp. 337–378. Amsterdam: Elsevier.Google Scholar
Chikawa, J. & Matsui, J. (1994). Growth and evaluation of bulk crystals. Handbook on semiconductors, completely revised edition, Vol. 3, edited by S. Mahajan, Chap. 1, pp. 1–71. Amsterdam: Elsevier.Google Scholar
Chikawa, J., Sato, F., Kawamura, T., Kuriyama, T., Yamashita, T. & Goto, N. (1986). A high-resolution video camera tube for live X-ray topography using synchrotron radiation. X-ray instrumentation for the Photon Factory: dynamic analyses of micro structures in matter, edited by S. Hosoya, Y. Iitaka & H. Hashizume, pp. 145–157. Tokyo: KTK Scientific Publishers.Google Scholar
Green, R. E. Jr (1977). Direct display of X-ray topographic images. Adv. X-ray Anal. 20, 221–235.Google Scholar
Heynes, G. D. (1977). Digital television; a glossary and bibliography. SMPTE J. 86, 6–9.Google Scholar
McMann, R. H., Kreinik, S., Moore, J. K., Kaiser, A. & Rossi, J. (1978). A digital noise reducer for encoded NTSC signals. SMPTE J. 87, 129–133.Google Scholar
McMaster, R. C., Photen, M. L. & Mitchell, J. P. (1967). The X-ray vidicon television image system. Mater. Eval. 25, 46–52.Google Scholar
Rose, A. (1948). The sensitivity performance of the human eye on an absolute scale. J. Opt. Soc. Am. 38, 196–208.Google Scholar
Rossi, J. P. (1978). Digital technique for reducing television noise. SMPTE J. 87, 134–140.Google Scholar
Rozgonyi, G. A., Haszlo, S. E. & Statile, J. L. (1970). Instantaneous video display of X-ray topographic images with resolving capabilities better than 15. Appl. Phys. Lett. 16, 443–446.Google Scholar
Sato, F., Maruyama, H., Goto, K., Fujimoto, I., Shidara, K., Kawamura, T. W., Hirai, T., Sakai, H. & Chikawa, J. (1993). Characteristics of a new high-sensitivity X-ray imaging tube for video topography. Jpn. J. Appl. Phys. 32, 2142–2146.Google Scholar