Section 11.4.3.3. Data-collection requirements

Z. Otwinowskia* and W. Minorb

a UT Southwestern Medical Center at Dallas, 5323 Harry Hines Boulevard, Dallas, TX 75390-9038, USA, and bDepartment of Molecular Physiology and Biological Physics, University of Virginia, 1300 Jefferson Park Avenue, Charlottesville, VA 22908, USA
Correspondence e-mail:  zbyszek@mix.swmed.edu

11.4.3.3. Data-collection requirements

| top | pdf |

The total oscillation range has to cover a sufficient number of spots to establish periodicity of the diffraction pattern in three dimensions. It is important that the oscillation range of each image is small enough so that the lunes (rings of spots, all from one reciprocal plane) are resolved. One should note that the requirement for lune separation is distinct from the requirement for spot separation. If lunes overlap, spots may have more than one index consistent with a particular position on the detector.

The autoindexing procedure described above is not dependent on prior knowledge of the crystal unit cell; however, for efficiency reasons, the search is restricted to a reasonable range of unit-cell dimensions, obtained, for example, from the requirement of spot separation. In DENZO, this default can be overridden by the keyword `longest vector', but the need to use this keyword is a sign of a problem that should be fixed. Either the defined spot size should be decreased or data should be recollected with the detector further away from the crystal.