X-ray-generating devices such as sealed tubes and rotating-anode generators produce intense beams of small cross section and are capable of giving severe radiation burns within a second or so of exposure. Great care is necessary when working close to the exit port of these devices.
Apertures in the housing enclosing the X-ray source should be covered by a shutter when the source is not being used. Interlocking devices should exist to prevent the emission of X-rays when:
|
Housings, shutters, shielded enclosures, and beam stops should be constructed such that the dose equivalent at any accessible point 0.05 m from their surface does not exceed 25 mSv for all practical operating conditions of the source.
Warning lights and illuminated signs should be fitted, interlocked such that they are lit when a shutter is open.