The selection of the X-ray tube anode is determined by several factors such as intensity, specimen fluorescence, and dispersion. The intensity of the characteristic line radiation varies among the target elements depending on the voltage and if a vacuum or He path is used. The recorded intensities also change abruptly at the absorption edges of the elements in the specimen. If a diffracted-beam monochromator or solid-state detector with narrow window centred on the characteristic line energy is used, the specimen fluorescence is eliminated (except for the element that is the same as the anode), and one tube can be used for all compositions. If the pattern has severe overlapping, the separation of the peaks can be increased with longer wavelengths, which increase the dispersion expressed as °θ Å−1 of d. Fig. 2.3.5.5
shows portions of diffractometer patterns of topaz in which the same d ranges were recorded with Cu Kα (a) and Cr Kα (b). The greater separation of the peaks is clearly advantageous in analysing the patterns.
![]() | Portion of diffractometer pattern of topaz showing effect of increasing dispersion on separation of peaks. (a) Cu Kα, (b) Cr Kα. |
Copper-anode tubes are most frequently used for powder work because of their high intensity and good dispersion. Chromium tubes are often used for specimens containing iron and other transition elements to avoid fluorescence, and for larger dispersion, but require a vacuum or helium path and the intensity is usually one-half or less than that of copper. Molybdenum tubes are often used for single-crystal analysis, but not often for powders because of the low dispersion.