International
Tables for Crystallography Volume C Mathematical, physical and chemical tables Edited by E. Prince © International Union of Crystallography 2006 |
International Tables for Crystallography (2006). Vol. C. ch. 2.6, p. 100
Section 2.6.1.5.2. Detectors
O. Glattera
|
In this field, we are facing the same situation as we met for X-ray sources. The detectors for small-angle scattering experiments are the same as or slightly modified from the detectors used in crystallography. Therefore, it is sufficient to give a short summary of the detectors in the following; further details are given in Chapter 7.1
. If we are not investigating the special cases of fully or partially oriented systems, we have to measure the dependence of the scattered intensity on the scattering angle, i.e. a one-dimensional function. This can be done with a standard gas-filled proportional counter that is operated in a sequential mode (Leopold, 1982
), i.e. a positioning device moves the receiving slit and the detector to the desired angular position and the radiation detector senses the scattered intensity at that position. In order to obtain the whole scattering curve, a series of different angles must be positioned sequentially and the intensity readings at every position must be recorded. The system has a very high dynamic range, but – as the intensities at different angles are measured at different times – the stability of the primary beam is of great importance.
This drawback is eliminated in the parallel detection mode with the use of position-sensitive detectors. Such systems are in most cases proportional counters with sophisticated and expensive read-out electronics that can evaluate on-line the accurate position where the pulses have been created by the incoming radiation.
Two-dimensional position-sensitive detectors are necessary for oriented systems, but they also have advantages in the case of non-oriented samples when circular chambers are used or when integration techniques in square detectors lead to a higher signal at large scattering angles.
The simplest and cheapest two-dimensional detector is still film, but films are not used very frequently in small-angle scattering experiments because of limited linearity and dynamic range, and fog intensity.
Koch (1988) reviews the one- and two-dimensional detectors actually used in synchrotron small-angle scattering experiments. For a general review of detectors, see Hendrix (1985
).
References


