International
Tables for Crystallography Volume C Mathematical, physical and chemical tables Edited by E. Prince © International Union of Crystallography 2006 |
International Tables for Crystallography (2006). Vol. C. ch. 2.6, p. 109
Section 2.6.2.6.2. Detector-response correction
R. Mayb
|
For geometrical reasons (e.g. sample absorption), and in the case of 2D detectors also for electronic reasons, the scattering curves cannot be measured with a sensitivity uniform over all the angular region. Therefore, the scattering curve has to be corrected by that of a sample with identical geometrical properties, but scattering the neutrons with the same probability into all angles (at least in the forward direction). As we have seen previously, such samples are vanadium and thin cells filled with light water. Again, water has the advantage of a much higher scattering cross section, and is less influenced by surface effects.
At large sample-to-detector distances (more than about 10 m), the scattering from water is not sufficiently strong to enable its use for correcting sample scattering curves obtained with the same settings. Experience shows that it is possible in this case to use a water scattering curve measured at a shorter sample-to-detector distance. This should be sufficiently large not to be influenced by the deviation of the (flat) detector surface from the spherical shape of the scattered waves and small enough so that the scattering intensity per detector element is still sufficient, for example about 3 m. It is necessary to know the intensity loss factor due to the different solid angles covered by the detector element and by the apparent source in both cases. This can be determined, for instance, by comparing the global scattering intensity of water on the whole detector for both conditions (after correction for the background scattering) or from the intensity shift of the same sample measured at both detector distances in a plot of the logarithm of the intensity versus Q.