Section 4.2.1.3. X-ray tubes

U. W. Arndta

4.2.1.3. X-ray tubes

| top | pdf |

The commonest source of X-rays is the high-vacuum, or Coolidge, X-ray tube, which may be either demountable and pumped continuously when in operation or permanently sealed after evacuation. The vacuum tube contains an electron gun that incorporates a thermionic cathode, which produces a well defined electron beam that is accelerated towards the anode or target, formerly often called the anticathode. In most X-ray tubes intended for crystallographic purposes, the anode is massive, i.e. its thickness is large compared with the range of the electrons; it is usually water-cooled and its surface is normal to the incident electron beam. Usually, it is desirable for the X-ray source to be small (between 25 μm and 1 mm square) and for the X-ray intensity from the tube to be the maximum possible for the amount of power that can be dissipated in the target. These objectives are best achieved by designing the electron gun to produce a line focus, that is the electron focus on the target face is approximately rectangular with the small dimension equal to the desired effective source size and the large dimension about 10 to 20 times larger. The focus is viewed at an angle between about 2 and 5° to the anode surface to produce an approximately square foreshortened effective source; and the X-ray windows are so positioned as to make these take-off angles possible. For some purposes, very fine line sources are required and windows may be provided to allow the focus to be viewed so as to foreshorten the line width. Higher power dissipation is possible in X-ray tubes in which the anode rotates: the line focus is now usually on the cylindrical surface of the anode with its long dimension parallel to the axis of rotation.

For focal-spot sizes down to about 100 μm, an electrostatic gun is adequate; this consists of a fine helical filament and a Wehnelt cathode, which produces a demagnified electron image of the filament on the anode. For most purposes, the Wehnelt cathode can be at the same potential as the filament but cleaner foci and adjustment of the focal spot size are possible when this electrode is negatively biased with respect to the filament. The filament is nearly always directly heated and made of tungsten. Lower filament temperatures, and smaller heating currents, could be achieved with activated heaters but the vacuum in high-power devices like X-ray tubes is rarely hard enough to permit their use since they are easily poisoned. However, Yao (1992[link]) has reported successful operation of a hot-pressed polycrystalline lanthanum hexaboride cathode in an otherwise unmodified RU-1000 rotating-target X-ray generator.

Very fine focus tubes, with foci in the range between 25 and 1 μm, require magnetic lenses. At one time, the all-electrostatic X-ray tube of Ehrenberg & Spear (1951[link]), which achieved foci between 20 and 80 μm, was very popular.

Sealed-off X-ray tubes for crystallographic use are nowadays made in the form of inserts containing a target of one of a range of standard metals to produce the desired characteristic radiation. A series of nominal focal-spot sizes, shown in Table 4.2.1.3[link], is commonly available. The insert is mounted inside a standard shield that is radiation- and shock-proof and that is fitted with X-ray shutters and filters and often also with a standardized track for mounting X-ray cameras. The water-cooled anode is normally at ground potential and the negative high voltage for the cathode, together with the filament supply, is brought in through a shielded shock-proof cable. The high voltage is nowadays generally of the constant-voltage type, that is, it is full-wave rectified and smoothed by means of solid-state rectifiers and capacitors housed in the high-voltage transformer tank, which also contains the filament transformer. The high tension and the tube current are frequently stabilized. Only the simplest X-ray generators now employ an alternating high tension that is rectified by the self-rectifying property of the X-ray tube itself.

Table 4.2.1.3| top | pdf |
Copper-target X-ray tubes and their loading

X-ray tubeAnode diameter (mm)Speedf1 × f2
(mm) (mm)
μLoading (kW)Recommended specific loading (kW mm−2)
r min−1mm s−1calc.recommended
Standard insert8 × 0.150.2951.00.80.67
8 × 0.40.3591.21.50.47
10 × 1.00.4251.82.00.20
12 × 2.00.4932.52.70.11
AEI-GX21896000280001 × 0.10.4251.41.212.0
   2 × 0.20.4253.953.28.0
   3 × 0.30.4257.35.25.8
   5 × 0.50.42515.615.06.0
AEI-GX1345745001080001 × 0.10.4252.72.727.0
Rigaku-RU200996000310001 × 0.10.4251.51.212.0
   2 × 0.20.4254.23.07.5
   3 × 0.30.4257.65.46.0
Rigaku-RU50040012502620010 × 0.50.35926.8306.0
Rigaku-RU100040025005245010 × 10.42560606.0
Rigaku-RU15002501000013100010 × 10.42596909.0
KFA-Jülich2501200015700014 × 1.40.4251731206.1

A demountable continuously pumped form of construction is nowadays adopted mainly for rotating-anode and other specialized X-ray tubes. The pumping system must be capable of maintaining a vacuum of better than 10−5 Torr: filament life is critically dependent upon the quality of the vacuum.

Rotating-anode tubes have been reviewed by Yoshimatsu & Kozaki (1977[link]). The first successful tube of this type that incorporated a vacuum shaft seal was described by Clay (1934[link]). Modern tubes mostly contain vacuum-oil-lubricated shaft seals of the type due to Wilson (1941[link]) and are based on, or are similar to, the rotating-anode tubes described by Taylor (1949[link], 1956[link]). In some tubes, successful use has been made of ferro-fluidic vacuum seals (see Bailey, 1978[link]). The main problems in the operation of rotating-anode tubes is the lifetime of the seals and of bearings that operate in vacuo. In successful tubes, e.g. those manufactured by Enraf, Rigaku-Denki, and Siemens, these lifetimes are about the same as the lifetime of the filament under good vacuum conditions, that is, of the order of 1000 h.

Phillips (1985[link]) has written a review article on stationary and rotating-anode X-ray tubes that contains many important practical details.

4.2.1.3.1. Power dissipation in the anode

| top | pdf |

The allowable power loading of X-ray tube targets is determined by the temperature of the target surface, which must remain below the melting point. Müller (1927[link], 1929[link], 1931[link]) first calculated the maximum loading both for stationary and for rotating anodes. His calculations were refined by Oosterkamp (1948a[link],b[link],c[link]) who considered, in particular, targets of finite thickness, and who also treated pulsed operation of the tube. For normal conditions, Oosterkamp's conclusions and those of Ishimura, Shiraiwa & Sawada (1957[link]) do not greatly differ from those of Müller, which are in adequate agreement with experimental observations.

For an elliptical focal spot with axes [f_1] and [f_2], Müller's formula for the maximum power dissipation on a stationary anode, assumed to be a water-cooled block of dimensions large compared with the focal-spot dimensions, can be written [W_{\rm stat}=2.063(T_M-T_0)Kf_1\mu(\,f_1,f_2),\eqno (4.2.1.12)]where K is the specific thermal conductivity of the target material in W mm−1, [T_M] is the maximum temperature at the centre of the focal spot on the target, that is, a temperature well below the melting point of the target material, and [T_0] is the temperature of the cold surface of the target, that is, of the cooling water. The function μ is shown in Fig. 4.2.1.4[link] . For copper, K is 400 W m−1 and, with [T_M-T_0] = 500 K, [W_{\rm stat}=425\mu f_1.\eqno (4.2.1.13)]For f2/f1 = 0.1, and μ = 0.425, this equation becomes [W_{\rm stat}=180 \, f_1.\eqno (4.2.1.14)]In these last two equations, f1 is in mm.

[Figure 4.2.1.4]

Figure 4.2.1.4| top | pdf |

The function μ in Müller's equation (equation 4.2.1.12[link]) as a function of the ratio of width to length of the focal spot.

For a rotating target, Müller found that the permissible power dissipation was given by [W_{\rm rot}=1.428\,K(T_M-T_0)\,f_1(\,f_2\rho Cv/2K)^{1/2},\eqno (4.2.1.15)]where f2 is the short dimension of the focus, assumed to be in the direction of motion of the target, v is the linear velocity, ρ is the density of the target material, and C is its specific heat.

For a copper target with f1 and f2 in mm and v in mm s−1, [W_{\rm rot}=26.4\,f_1(\,f_2v)^{1/2}.\eqno (4.2.1.16)]Equation (4.2.1.16)[link] shows that for very narrow focal spots rotating-anode tubes give useful improvements in permissible loading only if the surface speed is very high (see Table 4.2.1.3[link]). The reason is that with large foci on stationary anodes the isothermal surfaces in the target are planar; with fine foci, these surfaces become cylindrical and this already makes for very efficient cooling without the need for rotation. Rotating anodes are thus most useful for medium-size foci (200 to 500 μm) since for the larger focal spots it becomes very expensive to construct power supplies capable of supplying the permissible amount of power.

Table 4.2.1.3[link] shows the recommended loading for a number of commercially available X-ray tubes with copper targets, which will be seen to be in qualitative agreement with the calculations. Some of the discrepancy is due to the fact that the value of [K(T_M-T_0)] for the copper–chromium alloy targets used in actual X-ray tubes is appreciably lower than the value for pure copper used here. To a good approximation, the permissible loading for other targets can be derived by multiplying those in Table 4.2.1.3[link] by the factors shown in Table 4.2.1.4[link]. It is worth noting that the recommended loading of commercial stationary-target X-ray tubes has increased steadily in recent years. This is largely due to improvements in the water cooling of the back surface of the target by increasing the turbulence of the water and the effective surface area of the cooled surface.

Table 4.2.1.4| top | pdf |
Relative permissible loading for different target materials

CuCrFeCoMoAgW
1.00.90.60.91.21.01.2

In considering Table 4.2.1.3[link], it should be noted that the linear velocities of the highest-power X-ray-tube anode have already reached a speed that Yoshimatsu & Kozaki (1977[link]) consider the practical limit, which is set by the mechanical properties of engineering materials. It should also be noted that much higher specific loads can be achieved for true micro-focus tubes, e.g. 50 kW mm−2 for a 25 μm Ehrenberg & Spear tube and 1000 kW mm−2 for a tube with a 1 μm focus (Goldsztaub, 1947[link]; Cosslett & Nixon, 1951[link], 1960[link]).

Some tubes with focus spots of less than 10 μm utilize foil or needle targets. These targets and the heat dissipated in them have been discussed by Cosslett & Nixon (1960[link]). The dissipation is less than that in a massive target by a factor of about 3 for a foil and 10 for a needle, but, in view of the low absolute power, target movement and even water-cooling can be dispensed with.

References

First citation Bailey, R. L. (1978). The design and operation of magnetic liquid shaft seals. In Thermomechanics of magnetic fluids, edited by B. Berkovsky. London: Hemisphere.Google Scholar
First citation Clay, R. E. (1934). A 5 kW X-ray generator with a spinning target. Proc. Phys. Soc. London, 46, 703–712.Google Scholar
First citation Cosslett, V. E. & Nixon, W. C. (1951). X-ray shadow microscope. Nature (London), 168, 24–25.Google Scholar
First citation Cosslett, V. E. & Nixon, W. C. (1960). X-ray microscopy, pp. 217–222. Cambridge University Press.Google Scholar
First citation Ehrenberg, W. & Spear, W. E. (1951). An electrostatic focusing system and its application to a fine focus X-ray tube. Proc. Phys. Soc. London Sect. B, 64, 67–75.Google Scholar
First citation Goldsztaub, S. (1947). Tube à rayons X de grande brilliance à foyer ponctuel. C. R. Acad. Sci. 224, 458–459.Google Scholar
First citation Ishimura, T., Shiraiwa, Y. & Sawada, M. (1957). The input power limit of the cylindrical rotating anode of an X-ray tube. J. Phys. Soc. Jpn, 12, 1064–1070.Google Scholar
First citation Müller, A. (1927). Input limits of X-ray generators. Proc. R. Soc. London Ser. A, 117, 30–42.Google Scholar
First citation Müller, A. (1929). A spinning-target X-ray generator and its input limit. Proc. R. Soc. London Ser. A, 125, 507–516.Google Scholar
First citation Müller, A. (1931). Further estimates of the input limits of X-ray generators. Proc. R. Soc. London Ser. A, 132, 646–649.Google Scholar
First citation Oosterkamp, W. J. (1948a). The heat dissipation in the anode of an X-ray tube. Philips Res. 3, 49–59.Google Scholar
First citation Oosterkamp, W. J. (1948b). The heat dissipation in the anode of an X-ray tube. Philips Res. 3, 161–173.Google Scholar
First citation Oosterkamp, W. J. (1948c). The heat dissipation in the anode of an X-ray tube. Philips Res. 3, 303–317.Google Scholar
First citation Phillips, W. C. (1985). X-ray sources. Methods Enzymol. 114, 300–316.Google Scholar
First citation Taylor, A. (1949). A 5 kW crystallographic X-ray tube with a rotating anode. J. Sci. Instrum. 26, 225–229.Google Scholar
First citation Taylor, A. (1956). Improved demountable crystallographic rotating anode X-ray tube. Rev. Sci. Instrum. 27, 757–759.Google Scholar
First citation Wilson, R. R. (1941). A vacuum-tight sliding seal. Rev. Sci. Instrum. 12, 91–93.Google Scholar
First citation Yao, T. (1992). Lanthanum hexaboride filament in an X-ray generator of a laboratory EXAFS facility. Rev. Sci. Instrum. 63, 2103–2104.Google Scholar
First citation Yoshimatsu, M. & Kozaki, S. (1977). High-brilliance X-ray sources. X-ray optics, edited by H.-J. Queisser, Chap. 2. Berlin: Springer.Google Scholar








































to end of page
to top of page