International
Tables for
Crystallography
Volume C
Mathematical, physical and chemical tables
Edited by E. Prince

International Tables for Crystallography (2006). Vol. C. ch. 7.1, p. 625

Section 7.1.6.1.4. Spatial resolution

U. W. Arndtb

7.1.6.1.4. Spatial resolution

| top | pdf |

The spatial resolution of a PSD is determined by the number and size of resolution or picture elements (pixels) along the length or parallel to the edge of the detector. In most diffraction experiments, the size of the pattern can be scaled by altering the distance of the detector from the sample and what is important is the angular resolution of the detector when placed at a distance where it can `see' the entire pattern. We shall see below that linear PSD's can be made with up to 2000 pixels and that area detectors are mostly limited to fewer than 512 × 512 pixels. The sizes of pixels range from about 10 µm for semiconductor PSD's to about 1 mm for most gas-filled detectors.

The useful number of pixels of a detector is determined by its point-spread function (PSF). This is the relative response as a function of distance from the centre of a point image on the detector. PSF's are not necessarily radially symmetrical and may have to be specified in at least two directions at right angles, for example along and perpendicular to the lines of a television raster scan. The width of the PSF at the 50% level determines the amount of detail visible in a directly viewed image. The accuracy of intensity measurements may depend more critically upon the width of the PSF at a lower level, since a weak spot may be immeasurable when sitting on the `tail' of a very intense one. For various physical reasons, the PSF's of all PSD's, including X-ray film, have appreciable tails.

The spatial resolution of a detector is affected by parallax: when an X-ray beam is absorbed in a thick planar detector at an angle [\varphi] to the normal, the width of the resultant image is smeared out exponentially and its centroid is shifted by an amount [\sin\varphi/\mu]. For 8 keV X-rays incident at 45° on a xenon-filled counter, for example, this shift is about 4 mm for a filling pressure of 1 atm and 0.4 mm for a filling pressure of 10 atm. These figures illustrate the desirability of high-pressure xenon (Fig. 7.1.6.2[link] ) for gas-ionization detectors intended for wide-angle diffraction patterns.

[Figure 7.1.6.2]

Figure 7.1.6.2| top | pdf |

Absorption of 8 keV and 17 keV photons in argon and xenon as a function of pressure in atm × column length in mm.








































to end of page
to top of page