Section 7.2.4.1. Faraday cage

J. N. Chapmana

a Department of Physics and Astronomy, University of Glasgow, Glasgow G12 8QQ, Scotland

7.2.4.1. Faraday cage

| top | pdf |

The Faraday cage is the most convenient means of determining electron-beam currents. It consists of a small electrically isolated cage with a small hole in it through which electrons enter. Provided the hole subtends a sufficiently small solid angle and the inner surface of the cage is made of a material with a low back-scattering coefficient, the probability that any electrons will re-emerge is negligible. An electrometer is normally used to measure the charge that has entered the cage.

The main use of the Faraday cage is to calibrate other detectors when absolute electron intensities are required. It also serves a very important role when the total specimen exposure in an experiment must be kept below a critical value owing to the susceptibility of the specimen to radiation damage. As electron charge is being measured, the Faraday cage may be used with electrons of any energy.