Chapter 10.1. Introduction to cryocrystallography

H. Hopea*

aDepartment of Chemistry, University of California, Davis, One Shields Ave, Davis, CA 95616-5295, USA
Correspondence e-mail: hhope@ucdavis.edu

A general introduction to cryocrystallography is provided. The chapter covers crystal mounting, cooling of biocrystals, principles of cooling equipment and frost prevention. Operational considerations, such as dual-stream instruments, electrically heated nozzles, temperature calibration and transfer of crystals to difractometers, are also described.

Keywords: cryocrystallography; crystal mounting; data collection; frost prevention; ice formation; radiation damage.

10.1.1. Utility of low-temperature data collection

| top | pdf |

10.1.1.1. Prevention of radiation damage

| top | pdf |

Since about 1985, low-temperature methods in biocrystallography have moved from stumbling experimentation to mainstay production techniques. This would not have happened without good reason. A brief discussion of the advantages of data collection at cryogenic temperatures is given.

Biocrystals near room temperature are sensitive to X-rays and generally suffer radiation damage during data measurement. Often this damage is so rapid and severe that a number of different crystals are needed for a full data set. On occasion, damage is so rapid that data collection is impossible. Crystal decay is typically accompanied by changes in reflection profiles and cell dimensions, which alter the positions of diffraction maxima, exacerbating the problem of changing diffraction intensities. The use of more than one crystal almost always introduces inaccuracies. Intensities from a crystal near the end of its usable life will have decay errors. Individual samples of biocrystals frequently have measurable differences in structure; merging of data will result in an average of the structures encountered, with some loss of definition. Crystals cooled to near liquid-N2 temperature typically show a greatly reduced rate of radiation damage, often to the extent that it is no longer an issue of concern. The protection from radiation damage was noted early on. Petsko (1975)[link] observed numerous cases of this effect. A noteworthy example is the successful prevention of radiation damage to crystals of ribosome particles (Hope et al., 1989[link]).

Radiation damage appears to be related to the formation of free radicals. At sufficiently low temperature, two effects can influence the rate of damage: movement of the radicals is hampered, and the activation energy for reaction is not available. A revealing observation has been described by Hope (1990)[link]. A crystal that had been exposed to synchrotron radiation for many hours at 85 K showed no overt signs of radiation damage. However, while the crystal was being warmed toward room temperature, it suddenly turned black and curled up like a drying leaf. More commonly, crystals turn yellow under X-ray irradiation, and bubbles and cracks appear on warming. The rate of free-radical formation would be affected little by temperature, so that when sufficient mobility and activation energy become available, the stored radicals will react.

10.1.1.2. Mechanical stability of the crystal mount

| top | pdf |

The mechanical stability of samples is also of concern. Crystals mounted in capillaries and kept wet have a tendency to move, giving rise to difficulties with intensity measurements. A crystal at cryotemperature is rigidly attached to its mount; slippage is impossible.

10.1.1.3. Effect on resolution

| top | pdf |

The effects on radiation damage and mechanical stability are clear-cut, and provide the main reasons for using cryotechniques. Resolution can also be affected, but the connection between temperature and resolution is neither simple nor obvious. If low resolution is the result of rapid radiation damage, lowering the temperature can lead to much improved resolution. However, if low resolution is mainly caused by inexact replication from one unit cell to another, lowering the temperature may have little effect on resolution. If the mosaic spread in the crystal increases upon cooling, resolution may even deteriorate.

In a model proposed by Hope (1988[link]), a relationship between resolution r and temperature T is given by [r_{2} = r_{1}[(B_{0} + bT_{2})/(B_{0} + bT_{1})]^{1/2}.] Here [r_{1}] is the resolution at [T_{1}], [r_{2}] is the resolution at [T_{2}], [B_{0}] is the value of B at [T = 0] and b is a proportionality constant. The underlying assumption is that for any given temperature, the temperature factor [i.e. [\exp(-B\sin^{2}\theta/\lambda^{2})]] at the resolution limit has the same value; thus the effects of scattering factors and Lp factors are ignored. We see that if [B_{0}] is the predominant term, lowering T will not have much effect, whereas for small [B_{0}] (a relatively well ordered structure) the effect of T on r can be large. For example, if the room-temperature resolution is 1.5 Å, the resolution at 100 K can be around 1 Å, but if the room-temperature resolution is around 3 or 4 Å, little change can be expected. A qualitative assessment of these effects was clearly stated by Petsko (1975)[link].

10.1.2. Cooling of biocrystals

| top | pdf |

10.1.2.1. Physical chemistry of biocrystals

| top | pdf |

Crystals are normally brought from room temperature to the working, low temperature by relatively rapid cooling, either in a cold gas stream, or by immersion in a cryogen such as liquid nitrogen or liquid propane. One goal of the procedure is to avoid crystallization of any water present in the system, whether internal or external to the crystal. Ice formation depends on the formation of nuclei. Nuclei are formed either by homogenous nucleation, i.e. in bulk liquid, or by heterogeneous nucleation, i.e. at the surface of a phase other than the liquid. Although data pertaining to biocrystals are scarce, indications are that internal nucleation, whether homogenous or heterogeneous, is not common. Proteins that induce nucleation at mild supercooling are known, so presumably there exist regions in these proteins which help to prearrange water molecules so that they readily form ice nuclei. There are also proteins that hinder nucleation. At present there is no basis for predicting the outcome of cooling for any given protein crystal. Only in a statistical sense can one be reasonably confident that a given macromolecule will not promote the freezing of water.

Vali and coworkers (Götz et al., 1991[link]; Vali, 1995[link]) have provided a quantitative treatment of ice nucleation that can serve as a guideline. They observe that the absolute rate of formation of nuclei increases with the volume of water and with decreasing temperature. The probability p that a volume of water will freeze during a time span t is given by [p = J(T)Vt,] where J(T) is the nucleation rate at temperature T. Based on empirical data, J(T) is given by [J(T) = 6.8 \times 10^{-50} \exp[3.9(273 - T)],] where J is in m−3 s−1 and T is in K. Note that J(T) increases by a factor of 50 per K. As a practical limit, bulk water cannot be cooled below 233 K without freezing. However, given a sufficiently small volume and high cooling rate, it is possible to supercool water to form a glassy state that is at least kinetically stable. Stability requires a temperature below 140 K; at higher temperatures crystallization eventually takes place. For the cooling rates typically attained with small crystals (up to a few hundred K s−1) it seems impossible to avoid crystallization of water in the mother liquor adhering to a crystal, unless it is modified in some way. Once ice forms at the crystal surface, freezing may propagate through the entire crystal, effectively destroying it. Even if the crystal remains intact, diffraction from polycrystalline ice will render parts of any data set from that crystal useless. Because the probability of a nucleation event increases with time, it seems prudent to use a rapid cooling process. However, we note that the expression for J(T) is formulated for pure water and cannot be valid for all conditions; it is well established that a majority of biocrystals can be cooled below 140 K.

A consequence of the foregoing is that for prevention of ice growth one should first focus attention on the situation immediately outside the crystal, rather than on its interior. Two approaches have been shown to have merit: (a) modification of the solvent layer, and (b) removal of the solvent layer.

The goal of solvent modification is the prevention of ice formation in that layer. Commonly used modifiers (referred to as antifreezes or cryoprotectants) are water-soluble organic compounds of low molecular weight with good hydrogen-bonding properties; examples are glycerol, monomeric ethylene glycol and MPD (2-methyl-2,4-pentanediol). These compounds are used in sufficient concentration to suppress nucleation and thereby prevent ice formation. Typical concentrations are in the 15–30% range, depending on the compound and the original composition of the mother liquor. The required concentration must be determined by experiment. Some suitable starting points are given by Garman & Mitchell (1996)[link]. The modified solution is tested by cooling a small drop to the working temperature. If the drop remains clear, there is no ice formation.

It is important to keep in mind that any change in the properties of the medium surrounding the crystal will have consequences for its crystallographic stability. In order to protect the crystal, two fields should be considered: thermodynamics and kinetics.

For a crystal in equilibrium with its mother liquor, the chemical potential of each species will be the same inside the crystal and in the mother liquor. If the solution surrounding the crystal is altered by the addition of an antifreeze, the chemical potential µ of water (and other species) will change and the crystal will no longer be in equilibrium with its surrounding solution. The typical result is that μ(H2O, solution) decreases, so μ(H2O, crystal) > μ(H2O, solution) and there will be a thermodynamic drive to remove water from the crystal. The activation energy for water diffusion is low, so if the process is allowed to proceed, the end result is loss of water with likely deterioration in crystal quality (but see below). Considerations of this kind led Schreuder et al. (1988[link]) to develop procedures for solvent modification that would prevent destruction of the crystal. Although some success was reported, sufficient problems were encountered that the approach cannot be considered to be a general solution.

It is important to note that loss of water does not always lead to loss of crystal integrity. For example, Esnouf et al. (1998[link]) and Fu et al. (1999[link]) have shown that controlled dehydration can result in substantially improved resolution. In addition, antifreeze concentrations substantially higher than those needed to suppress ice formation (Mitchell & Garman, 1994[link]) can preserve low mosaic spread. These phenomena may be connected.

In earlier work, Travers & Douzou (1970[link]) emphasized the importance of keeping the dielectric constant unchanged when modifying the mother liquor. Petsko (1975[link]) made observations that support the significance of this approach and, based on systematic studies, also showed that keeping μ(H+) constant is of great importance. Hui Bon Hoa & Douzou (1973[link]) and Douzou et al. (1975[link]) have presented tables of solvent compositions that facilitate the preparation of successful cryoprotective solutions. It should be noted that a significant aim in Petsko's work was to keep the solvent liquid, so as to permit manipulation of enzyme substrates. Studies of enzyme kinetics are much more demanding than the rapid cooling to about 100 K that is of primary interest here.

In most cases it is only necessary to consider kinetic effects, i.e., how long it takes before the crystal itself begins to change. When a crystal in a drop of its original mother liquor is dipped into a drop of modified mother liquor, diffusion begins immediately. The speed of propagation in the liquid phase can be estimated from a standard equation for the mean-square travel distance of a diffusing species, [\overline{x^{2}} = 2Dt,] where D is the diffusion coefficient and t is the the time. Typical room-temperature values for D for antifreeze molecules in water are around 10-9 m2 s−1. Thus, a root-mean-square travel distance of 0.1 mm requires about 5 s. For a solvent layer about 0.1–0.2 mm thick, a contact time of 5–20 s will then provide a sufficient level of modification to prevent freezing, while the risk of crystal damage is small. It is often important to stop any ongoing process as soon as protection from freezing has been attained. This can conveniently be done by immersion in liquid N2.

10.1.2.2. Internal ice or phase transition

| top | pdf |

If there are good indications that ice formation does start internally, or that a destructive phase transition takes place, an attempt can be made to modify the internal water structure. An important consideration of Petsko (1975[link]) was never to allow large deviations from equilibrium. This can be accomplished by a slow, gradual change in μ(H2O, solution), allowing enough time for the crystal to re-establish equilibrium. A number of successful experiments were reported.

10.1.2.3. Removal of the solvent layer

| top | pdf |

Because of their tendency for rapid loss of internal solvent, dried biocrystals rarely survive exposure to the atmosphere. A solution to this problem was described by Hope (1988[link]). The solvent is removed while the crystal is submerged in a hydrocarbon oil. After the liquid has been removed, a small drop of oil is allowed to encapsulate the crystal, allowing it to tolerate brief exposure to air. Even under such mild conditions, some crystals still lose water and suffer damage. A remedy for this is to keep the oil saturated with water. One disadvantage of the oil technique is the tendency to carry along too much oil, which can cause excessive background scattering. One advantage is that absorption can become nearly isotropic. The most commonly used oil is Infineum Parabar 10312, formerly known as Exxon Paratone-8277 or Paratone-N.

10.1.2.4. Cooling rates

| top | pdf |

The time dependence of nucleation probability suggests that faster is safer. Although no systematic data are available, it is commonly assumed that crystal cooling should be as rapid as possible. Studies related to cryopreservation of biological samples for electron microscopy provide a number of measurements of cooling rates in various coolants, but it is difficult to extract information directly relevant to cryocrystallography. From a practical point of view, the coolants to be considered are liquid N2 and liquid propane (and, to a lesser extent, liquid ethane). Thermal conductivities for small-molecule compounds in liquid form tend to be of similar magnitude – around 1.5 × 10-5 W m−1 K−1. N2 boils at 77 K; propane remains liquid between 83 and 228 K. It is often thought that a gas bubble that can form around an object dipped in liquid N2 makes it less effective as a coolant than liquid propane, which is much less likely to form bubbles. However, from model calculations, Bald (1984[link]) suggested that the gas insulation problem in liquid N2 would not be significant in the cooling of small objects of low thermal conductivity, because there is not enough heat transport to the surface to maintain the gas layer. He also concluded that liquid N2 could potentially yield the highest cooling rate among commonly used coolants. But in a review of plunge-cooling methods, Ryan (1992[link]) gives preference to liquid ethane. Walker et al. (1998[link]) measured the cooling rates in N2 gas (100 K), liquid N2 (77 K) and liquid propane (100 K) of a bare thermocouple and of a thermocouple coated with RTV silicone cement. The thermocouples were made from 0.125-mm wire and the coating was about 0.20–0.25 mm thick. With the gas stream, cooling of the centres of the samples from 295 K to 140 K took 0.8 and 2 s, respectively; with liquid N2 the times were 0.15 and 0.6 s, and with liquid propane they were 0.15–0.18 and 1.2 s (time reproducibility is to within ±10%). Given the simplicity of liquid-N2 immersion, there seems little reason to choose the more complicated and more hazardous liquid-propane technique.

10.1.3. Principles of cooling equipment

| top | pdf |

There are many ways to construct a low-temperature apparatus based on the cold-stream principle that functions well, but they are all made according to a small number of basic principles.

All gas-stream crystal-cooling devices must have three essential components: (a) a cold gas supply, (b) a system of cold gas delivery to the crystal, and (c) a system for frost prevention at the crystal site.

10.1.3.1. Cold gas supply

| top | pdf |

Two methods are commonly used: generation of gas by boiling liquid N2 with an electrical heater, and cooling of a gas stream in a liquid-N2 heat exchanger.

Because precise voltage and current control are easily realized, the boiler method has the advantage of providing very accurate control of the flow rate with minimal effort. Precise control of the flow rate is typically not attained when the rate is controlled with standard gas-flow regulators, because they control volume, not mass.

In addition to control of the flow rate, precise control of the temperature requires exceptional insulation for the cold stream. The longer the stream path, the higher the requirements for insulation. As a rule, temperature rise during transfer should not exceed 15 K at a flow rate of 0.2 mol N2 min−1; preferably, it should be significantly lower. Higher cooling loss leads to excessive coolant consumption and to instability caused by changes in ambient temperature. High flow rates also tend to cause undesirable cooling of diffractometer parts. No commercially offered device should be accepted if it does not meet the criterion given above.

Appropriate insulation can be readily attained either with silvered-glass Dewar tubing or with stainless-steel vacuum tubing. Glass has the advantage of being available from local glassblowing shops; it generally provides excellent insulation. The main disadvantages are fragility and a rigid form that makes accurate positioning of the cold stream difficult. Stainless steel can provide superb insulation, given an experienced manufacturer; unsatisfactory insulation is quite common. A major advantage is the availability of flexible transfer lines that greatly simplify the positioning of the cold stream relative to the diffractometer.

10.1.3.2. Frost prevention

| top | pdf |

Three areas must be kept frost-free: the crystal, the crystal mount and the delivery end of the transfer tube. The first successful solution to this problem was the dual-stream design of Post et al. (1951[link]). It provides for a cold stream surrounded by a concentric warm stream. If the warm stream is sufficiently dry, this will prevent frost around the outlet. The crystal will remain frost-free only if mixing of the two flows occurs downstream from the crystal. For a stream aligned with the axis of the goniometer head, an additional shield is needed to keep the goniometer head frost-free.

10.1.4. Operational considerations

| top | pdf |

10.1.4.1. Dual-stream instruments

| top | pdf |

Fig. 10.1.4.1[link] shows a schematic drawing of the region around the crystal in a traditional dual-stream apparatus, first described by Post et al. (1951[link]). The device provides for a cold stream surrounded by a concentric warm stream. The diameter of the cold stream is typically around 7 mm with a shield stream of 2–3 mm. The two streams flow parallel to the axis of the crystal mount. In a properly functioning apparatus, the warm stream supplies enough heat to keep the tip of the tube carrying the cold stream above the dew point. It is important that the streams do not mix, or the crystal temperature will not be stable. This is achieved by careful balancing of flow rates to minimize turbulence. (Absence of turbulence can be judged by the shape of the shadow of the cold stream in a parallel beam of bright light.) In a laminar cold stream, the crystal is well protected and no unusual precautions are needed. The region of constant, minimum temperature will typically have a diameter of about 3 mm. Turbulent flow will result in no constant-temperature region, so it is important to verify the stream quality.

[Figure 10.1.4.1]

Figure 10.1.4.1| top | pdf |

Schematic drawing of a dual-stream setup with the streams parallel to the diffractometer φ axis. The top part represents the outlet end of the stream delivery device. A represents the outline of the warm shield stream and B represents the interface between the cold stream and the warm stream. The goniometer head (not shown) is protected by a shield.

The cold stream has sufficient heat capacity to cool down the goniometer head, and sometimes other adjacent equipment parts as well. A simple solution consists of an aluminium cone equipped with a heating coil on the back. A shield that functions well has been described by Bellamy et al. (1994[link]).

Fig. 10.1.4.2[link] illustrates a situation where the stream direction deviates substantially from the head-on direction in Fig. 10.1.4.1.[link] An angle of 35–55° will give good results. An advantage of an angled delivery is that the goniometer head will not be touched by the cold stream, therefore the heated stream deflector is not needed, resulting in simplified installation and operation.

[Figure 10.1.4.2]

Figure 10.1.4.2| top | pdf |

Schematic drawing of a dual-stream setup with the streams angled relative to the diffractometer φ axis. A and B are the same as in Fig. 10.1.4.1[link]. The cold stream misses the goniometer head, so no shield is required.

Analysis of the dual-stream apparatus reveals a twofold function of the outer stream: it keeps the nozzle frost-free and it supplies heat to the mounting pin. Protection of the crystal is, in reality, already provided by the laminar cold stream. The nozzle can be kept frost-free simply with an electric heater. Ice formation on the crystal mount can be easily suppressed by appropriate design of the mounting pin and mounting fibre, and attention to their interaction with the cold stream. A successful solution is sketched in Figs. 10.1.4.3[link] and 10.1.4.4[link].

[Figure 10.1.4.3]

Figure 10.1.4.3| top | pdf |

Schematic drawing of a single-stream setup with the stream parallel to the diffractometer φ axis. A represents the outline of the cold stream. The tip of the outlet end is heated above the dew point with a heating coil B. The goniometer head (not shown) is protected by a shield.

[Figure 10.1.4.4]

Figure 10.1.4.4| top | pdf |

Schematic drawing of a single-stream setup with the stream angled relative to the diffractometer φ axis. A represents the outline of the cold stream. At B the crystal mounting pin protrudes 1–2 mm into the cold stream. This prevents frost from forming on the mounting fibre. The tip of the outlet end is heated above the dew point with a heating coil C. The cold stream misses the goniometer head, so no shield is required. In general, the simplest operation is attained with a setup similar to that shown here.

10.1.4.2. Electrically heated nozzle

| top | pdf |

Fig. 10.1.4.3[link] shows the functional equivalent of Fig. 10.1.4.1[link]. Instead of the warm stream, an electrical heating element is used to keep the tip of the delivery tube ice free. An actual construction will usually consist of a nozzle that can be attached to the delivery tube. The heating element is made from standard resistance wire (e.g. Nichrome). About 5 W will usually be enough to prevent frost or condensation. The head-on direction results in reliable frost protection for the crystal, but the goniometer head must be equipped with a heated stream deflector.

Fig. 10.1.4.4[link] shows the functional equivalent of Fig. 10.1.4.2[link]. This arrangement leads to the simplest design, though some precautions are needed. The mounting pin itself supplies the heat needed to prevent ice formation on the pin. The cone-shaped tip of the pin (full cone angle about 20°) should be smooth in order to prevent turbulence. About 1–2 mm (but not more) of the tip must protrude into the cold stream (see region A in Fig. 10.1.4.4[link]). If too much of the pin is in the cold stream, the rest of the pin can become too cold and ice up. If the pin is too far out of the stream it will also fail, because glass or other insulating mounting materials will invariably collect frost at the cold/warm interface. As frost prevention depends on heat conducted from the rest of the pin, it must be made from copper. This design has been extensively tested and has been used for many years for the collection of a large number of data sets. Results are uniformly good, with simple operation and reliable frost prevention even in high humidity.

Omission of the warm stream results in significant design simplification. The entire apparatus for production of the outer stream is left out, resulting in real savings in manufacture, operation and maintenance. There is no obvious disadvantage. Ice protection is as good as with the dual stream. The main cost to the user is in the requirement that the mounting system be constructed within somewhat narrower limits. The fact that operator errors tend to become apparent through frost formation can actually be an advantage. With a dual-stream device, an improperly positioned cold stream or an improperly prepared crystal mount may not produce overt signs, even though the crystal temperature is ill-defined.

In all configurations shown, correct positioning of the cold stream is essential. The centre of the stream should not miss the centre of the diffractometer (and hence the crystal) by more than 0.5 mm.

10.1.4.3. Temperature calibration

| top | pdf |

Measurement of the temperature at the crystal site with thermocouples or other devices that require attached leads is very difficult, mainly because of heat conduction along the leads. The preferred method of calibration makes use of the known temperature of a phase transition of a crystal in the normal data-collection position. KH2PO4 (often referred to as KDP) has a sharp transition at 123 K from tetragonal to orthorhombic, and is commonly used. Another possibility is KH2AsO4, which has a corresponding phase transition at 95 K.

Two readout temperatures suffice, one at room temperature and one at the phase transition. The difference between readout temperature and crystal-site temperature can be assumed to vary linearly with T, so interpolation or extrapolation is simple.

10.1.4.4. Transfer of the crystal to the diffractometer

| top | pdf |

Inspection of Figs. 10.1.4.1[link] [link] [link]–10.1.4.4[link] reveals that the mounting of a crystal on a mounting pin via the traditional placement of the pin in the hole of a standard goniometer head is not simple, because the cooling nozzle is in the way. The solution to the problem is a design that allows side entry. Two methods are in use. One depends on a side-entry slot on a modified goniometer head; the slot is equipped with a spring-loaded catch that allows a very smooth, but stable, catch of the pin. The other method relies on a magnetic platform on the goniometer head and a corresponding magnetic base on the mounting pin.

The use of liquid-N2 cooling and side entry, and the requirement of reproducible knowledge of crystal temperature, led to the development of a set of tools for crystal mounting, as described by Parkin & Hope (1998[link]). The tools include special transfer tongs used for moving crystals from liquid N2 to the goniometer head. The temperature of the crystal is maintained by the heat capacity and low heat conductance of the tongs. The operation is independent of the orientation of the goniometer head, since there is no liquid to contain.

10.1.5. Concluding note

| top | pdf |

With correctly functioning low-temperature equipment and appropriate techniques, a crystal can be maintained frost-free for the duration of a data-collection run. Formation of frost on the crystal indicates malfunction of the equipment, or operator error. The most likely cause is operator error, but faulty equipment cannot be ruled out. The techniques described here have been used for collecting thousands of data sets from ice-free crystals and crystal mounts. There is no reason to accept frost problems as an unavoidable part of cryocrystallography.

References

First citation Bald, W. B. (1984). The relative efficiency of cryogenic fluids used in the rapid quench cooling of biological samples. J. Microsc. 134, 261–270.Google Scholar
First citation Bellamy, H. D., Phizackerley, R. P., Soltis, S. M. & Hope, H. (1994). An open-flow cryogenic cooler for single-crystal diffraction experiments. J. Appl. Cryst. 27, 967–970.Google Scholar
First citation Douzou, P., Hui Bon Hoa, G. & Petsko, G. A. (1975). Protein crystallography at sub-zero temperatures: lysozyme–substrate complexes in cooled mixed solutions. J. Mol. Biol. 96, 367–380.Google Scholar
First citation Esnouf, R. M., Ren, J., Garman, E. F., Somers, D. O'N., Ross, C. K., Jones, E. Y., Stammers, D. K. & Stuart, D. I. (1998). Continuous and discontinuous changes in the unit cell of HIV-1 reverse transcriptase crystals on dehydration. Acta Cryst. D54, 938–953.Google Scholar
First citation Fu, Z.-Q., Du Bois, G. C., Song, S. P., Harrison, R. W. & Weber, I. T. (1999). Improving the diffraction quality of MTCP-1 crystals by post-crystallization soaking. Acta Cryst. D55, 5–7.Google Scholar
First citation Garman, E. F. & Mitchell, E. P. (1996). Glycerol concentrations required for cryoprotection of 50 typical protein crystallization solutions. J. Appl. Cryst. 29, 584–587.Google Scholar
First citation Götz, G., Mészáros, E. & Vali, G. (1991). Atmospheric particles and nuclei, p. 142. Budapest: Akadémiai Kiadó.Google Scholar
First citation Hope, H.(1988). Cryocrystallography of biological macromolecules: a generally applicable method. Acta Cryst. B44, 22–26.Google Scholar
First citation Hope, H. (1990). Cryocrystallography of biological macromolecules at ultra-low temperature. Annu. Rev. Biophys. Biophys. Chem. 19, 107–126.Google Scholar
First citation Hope, H., Frolow, F., von Böhlen, K., Makowski, I., Kratky, C., Halfon, Y., Danz, H., Webster, P., Bartels, K. S., Wittmann, H. G. & Yonath, A. (1989). Cryocrystallography of ribosomal particles. Acta Cryst. B45, 190–199.Google Scholar
First citation Hui Bon Hoa, G. & Douzou, P. (1973). Ionic strength and protein activity of supercooled solutions used in experiments with enzyme systems. J. Biol. Chem. 248, 4649–4654.Google Scholar
First citation Mitchell, E. P. & Garman, E. F. (1994). Flash freezing of protein crystals: investigation of mosaic spread and diffraction limit with variation of cryoprotectant concentration. J. Appl. Cryst. 27, 1070–1074.Google Scholar
First citation Parkin, S. & Hope, H. (1998). Macromolecular cryocrystallography: cooling, mounting, storage and transportation of crystals. J. Appl. Cryst. 31, 945–953. Google Scholar
First citation Petsko, G. A. (1975). Protein crystallography at sub-zero temperatures: cryoprotective mother liquors for protein crystals. J. Mol. Biol. 96, 381–392.Google Scholar
First citation Post, B., Schwartz, R. S. & Fankuchen, I. (1951). An improved device for X-ray diffraction studies at low temperatures. Rev. Sci. Instrum. 22, 218–220. Google Scholar
First citation Ryan, K. P. (1992). Cryofixation of tissues for electron microscopy: a review of plunge cooling methods. Scanning Microsc. 6, 715–743.Google Scholar
First citation Schreuder, H. A., Groendijk, H., van der Lan, J. M. & Wierenga, R. K. (1988). The transfer of protein crystals from their original mother liquor to a solution with a completely different precipitant. J. Appl. Cryst. 21, 426–429.Google Scholar
First citation Travers, F. & Douzou, P. (1970). Dielectric constants of alcoholic-water mixtures at low temperature. J. Phys. Chem. 74, 2243–2244.Google Scholar
First citation Vali, G. (1995). In Biological ice nucleation and its applications, edited by R. E. Lee Jr, G. J. Warren & L. V. Gusta, pp. 1–28. St Paul: APS Press. Google Scholar
First citation Walker, L. J., Moreno, P. O. & Hope, H. (1998). Cryocrystallography: effect of cooling medium on sample cooling rate. J. Appl. Cryst. 31, 954–956.Google Scholar