Section 10.2.5. Transfer and storage
a
Department of Biochemistry, Chandler Medical Center, University of Kentucky, 800 Rose Street, Lexington, KY 40536-0298, USA |
Crystals flash cooled in a liquid cryogen must be placed for data collection in the cold gas stream of a cryostat without any substantial warming. One common transfer method (Rodgers, 1994, 1997) is shown in Fig. 10.2.5.1. Once the loop assembly has been plunged into the Dewar of liquid nitrogen, it is inserted into a small plastic vial of the type normally used for cryogenic storage, ensuring that the sample remains below the liquid surface during the operation. There is then sufficient liquid nitrogen in the vial to keep the sample cold as it is transferred to the cryostat gas stream. Again, the magnetic mounting system is used to reduce the time required for transfer. For X-ray cameras with vertical spindles, as shown in Fig. 10.2.5.1, some means of pointing the magnetic mount downward is required to prevent the nitrogen from spilling out of the vial. The goniometer illustrated has a detachable arc extension (Engel et al., 1996; Litt et al., 1998) that provides this capability. When the loop assembly is attached to the magnet, the vial is quickly withdrawn, exposing the crystal to the gas stream. The arc slide can then be returned to the normal position and the arc extension removed.
Another device (Mancia et al., 1995) for achieving the correct transfer geometry is shown in Fig. 10.2.5.2. This `flipper' mechanism can be extended to permit transfer of the crystal. The device is then rotated about the hinge to reorient the loop assembly for data collection. The hinge is positioned so that rotation does not translate the crystal, keeping it in the cold stream during reorientation.
When cooling in other liquid cryogens such as propane, the same cryovial transfer system is used. Flash cooling in cryovials (Fig. 10.2.4.2c) permits direct transfer using the magnetic mounting system. Usually, the liquid cryogen has been solidified in the vial, and it is allowed to melt at least partially before placing the crystal on the goniometer. Any remaining solid then melts and drips away (although it is often necessary to remove the last drop on the crystal with filter paper). When cooling in a larger volume of cryogen (Fig. 10.2.4.2b), the crystal can be `hopped' rapidly from the cooling cryogen to the surrounding vat of liquid nitrogen. A drop of cryogen transfers with the crystal, keeping it from warming. The loop assembly can then be placed in a cryovial and transferred to the goniometer.
Another device that does not use cryovials has been introduced (Parkin & Hope, 1998) to facilitate transfer from liquid nitrogen. The device consists of a split metal cup attached to handles that allow the cup to be opened and closed. When closed, the two halves of the metal cup form a cavity that can accommodate and grasp the loop assembly. As shown in Fig. 10.2.5.3, the loop assembly is inserted after first cooling the tongs in liquid nitrogen. The thermal mass of the tongs prevents warming as the crystal is then placed on the goniometer. The tongs are opened and removed to expose the crystal to the gas stream.
Any of these transfer procedures can be reversed in order to return the loop assembly to liquid nitrogen without thawing the crystal. The assembly and cryovial can then be placed in a Dewar designed for long-term storage. Some opening should be present in the loop-assembly bases, or the cryovials should be notched, to allow free movement of liquid nitrogen. The vials are conveniently held and organized using aluminium canes, which take up to five samples and have tabs that hold the loop assemblies in place. For even more secure long-term storage, loop assemblies with threaded bases that screw into the cryovials are available.
The ability to store samples for long periods of time permits a number of crystals to be flash cooled under consistent conditions, which can be important for maintaining isomorphism, and crystals can also be stockpiled for later data collection at a synchrotron X-ray facility. In fact, crystals should be prescreened for quality in the laboratory before synchrotron data collection to make efficient use of time on the beam line. Finally, crystals that degrade in growth or harvest solutions, or that contain macromolecules in unstable or transient states, can be conveniently preserved by flash cooling and storage in liquid nitrogen.
References
Abdel-Meguid, S. S., Jeruzalmi, D. & Sanderson, M. R. (1996). Crystallographic methods and protocols, edited by C. Jones, B. Mulloy & M. R. Sanderson, pp. 55–87. New Jersey: Humana Press.Google ScholarEngel, C., Wierenga, R. & Tucker, P. A. (1996). A removable arc for mounting and recovering flash-cooled crystals. J. Appl. Cryst. 29, 208–210.Google Scholar
Garman, E. F. & Schneider, T. R. (1997). Macromolecular cryocrystallography. J. Appl. Cryst. 30, 211–237.Google Scholar
Litt, A., Arnez, J. G., Klaholz, B. P., Mitschler, A. & Moras, D. (1998). A eucentric goniometer head sliding on an extended removable arc modified for use in cryocrystallography. J. Appl. Cryst. 31, 638–640.Google Scholar
Mancia, F., Oubridge, C., Hellon, C., Woollard, T., Groves, J. & Nagai, K. (1995). A novel device for the recovery of frozen crystals. J. Appl. Cryst. 28, 224–225.Google Scholar
Parkin, S. & Hope, H. (1998). Macromolecular cryocrystallography: cooling, mounting, storage and transportation of crystals. J. Appl. Cryst. 31, 945–953.Google Scholar
Rodgers, D. W. (1994). Cryocrystallography. Structure, 2, 1135–1140.Google Scholar
Rodgers, D. W. (1997). Practical cryocrystallography. Methods Enzymol. 276, 183–203.Google Scholar