Automated robotic crystal handling
Hope, H. and Parkin, S.  International Tables for Crystallography (2012). Vol. F, Section 10.1.4.5, p. 247 [ doi:10.1107/97809553602060000827 ]
Automated robotic crystal handling 10.1.4.5. Automated robotic crystal handling The era of structural genomics has necessitated a move to high throughput at high-intensity synchrotron sources. To meet this goal, a number of robotic crystal-handling devices have been developed. These in turn have had the beneficial effect of increased standardization ...

Transfer of the crystal to the diffractometer
Hope, H. and Parkin, S.  International Tables for Crystallography (2012). Vol. F, Section 10.1.4.4, p. 247 [ doi:10.1107/97809553602060000827 ]
... a set of tools for crystal mounting as described by Parkin & Hope (1998). The tools include special transfer tongs used ... goniometer head because there is no liquid to contain. References Parkin, S. & Hope, H. (1998). Macromolecular cryocrystallography: cooling, mounting, storage ...

Temperature calibration
Hope, H. and Parkin, S.  International Tables for Crystallography (2012). Vol. F, Section 10.1.4.3, pp. 246-247 [ doi:10.1107/97809553602060000827 ]
Temperature calibration 10.1.4.3. Temperature calibration Measurement of the temperature at the crystal site with sensing devices that require attached leads is very difficult, mainly because of heat conduction along the leads. It is usually necessary to loop the leads into the delivery nozzle. Both Si diodes and Pt resistance sensors have ...

Electrically heated nozzle
Hope, H. and Parkin, S.  International Tables for Crystallography (2012). Vol. F, Section 10.1.4.2, pp. 245-246 [ doi:10.1107/97809553602060000827 ]
Electrically heated nozzle 10.1.4.2. Electrically heated nozzle Fig. 10.1.4.3 shows the functional equivalent of Fig. 10.1.4.1. Instead of the warm stream, an electrical heating element is used to keep the tip of the delivery tube ice free. An actual construction will usually consist of a nozzle that can be attached to ...

Dual-stream instruments
Hope, H. and Parkin, S.  International Tables for Crystallography (2012). Vol. F, Section 10.1.4.1, p. 245 [ doi:10.1107/97809553602060000827 ]
... shown here. References Bellamy, H. D., Phizackerley, R. P., Soltis, S. M. & Hope, H. (1994). An open-flow cryogenic cooler ... J. Appl. Cryst. 27, 967-970. Post, B., Schwartz, R. S. & Fankuchen, I. (1951). An improved device for X-ray ...

Operational considerations
Hope, H. and Parkin, S.  International Tables for Crystallography (2012). Vol. F, Section 10.1.4, pp. 245-247 [ doi:10.1107/97809553602060000827 ]
... a set of tools for crystal mounting as described by Parkin & Hope (1998). The tools include special transfer tongs used ... pin geometry. References Bellamy, H. D., Phizackerley, R. P., Soltis, S. M. & Hope, H. (1994). An open-flow cryogenic cooler ... single-crystal diffraction experiments. J. Appl. Cryst. 27, 967-970. Parkin, S. & Hope, H. (1998). Macromolecular cryocrystallography: cooling, mounting, ...