Section 11.2.6.9. Systematic errors in profile-fitted intensities
aMRC Laboratory of Molecular Biology, Hills Road, Cambridge CB2 2QH, England |
The fundamental assumption in profile fitting is that the standard profiles accurately reflect the true profile of the reflection being integrated. Errors in the standard profile will result in systematic errors in the profile-fitted intensities. While these errors will often be small compared to the random (Poissonian) error for weak reflections, this is not necessarily the case for strong reflections, as the systematic error is typically a small percentage of the total intensity. Because the standard profiles are derived from the summation of many contributing reflections, small positional errors in spot prediction will lead to a broadening of the standard profile relative to the profile of an individual spot. The same broadening can occur because of the finite sampling interval in the image, which means that a predicted spot position can lie up to half a pixel away from the centre of the measurement box. This error can be minimized by interpolating the pixel values in the image onto a grid which is centred exactly on the predicted position, but the interpolation step itself will inevitably distort the reflection profile. In spite of these difficulties, providing adequate care is taken to determine the crystal and detector parameters accurately (as mentioned in Section 11.2.2), so that the spot positions are predicted to within a small fraction of the overall spot width, there is no suggestion (from merging statistics at least) for significant systematic error, even in the stronger intensities.