Section 7.2.2. CCD detector assembly

M. W. Tate,a* E. F. Eikenberryb and S. M. Grunera

aDepartment of Physics, 162 Clark Hall, Cornell University, Ithaca, NY 14853-2501, USA, and  bSwiss Light Source, Paul Scherrer Institut, 5232 Villigen PSI, Switzerland
Correspondence e-mail:  mwt5@.cornell.edu

7.2.2. CCD detector assembly

| top | pdf |

Any practical detector requires compromises in the choice of components to optimize those aspects most important to the diffraction problem at hand. The optimization of one detector characteristic often adversely affects other characteristics, so that it is often difficult to identify the `best' component. Considerations are given below to aid in making judicious choices.

Most fundamentally, a viable X-ray detector must have good quantum efficiency (see Section 7.1.1[link] in Chapter 7.1). Necessary, but not sufficient, are a high stopping power for X-rays and a large average signal per X-ray recorded in the CCD. The input signal passes through a sequence of stages and is transformed several times in the process. The statistics of this process are governed by the quantized nature of the signal, whether it be the initial X-ray photon, the visible photons produced in the phosphor, intermediate photoelectrons in intensifiers or the integrated charge in the CCD. To maintain a high detective quantum efficiency (DQE), the associated number of quanta per X-ray must be kept well above unity at each stage in the `quantum chain'. There are several approaches that meet this criterion.

Crystallography applications generally benefit from large detective areas, whereas typical scientific CCDs are quite modest in size (circa 25 × 25 mm). The usual solution is to use fibre-optic tapers (which are more efficient demagnifiers than lenses; see Deckman & Gruner, 1986[link]) to optically reduce a diffraction image excited in a larger phosphor screen. However, since optical image reduction is inherently inefficient, the reduction ratio is usually limited to about 4:1 before the number of visible photons per X-ray transmitted to the CCD becomes unacceptably low. Higher reduction ratios require image intensification before reduction (Moy, 1994[link]; Naday et al., 1995[link]; Tate et al., 1997[link]) or there will be an unacceptable loss in DQE. Intensification after reduction can result in the same average recorded signal per X-ray. However, there is a significant probability that no quanta at all make it through the chain for many of the incident X-rays, thereby lowering the number of X-rays actually `counted'.

Properties of the individual components affect other important detector characteristics as well. Below is a summary of important parameters for each of the components. Performance variations of CCD detectors from different vendors can most often be traced to the quality of the phosphor screen and the calibrations that are applied to the detector.

Phosphors. Although there are a bewildering variety of phosphor types, only a few are typically used with X-ray detectors (Shepherd et al., 1995[link]). A dense, high atomic number material is necessary to make the thin screens required for good spatial resolution while maintaining high X-ray stopping power. Gd2O2S:Eu offers high light output (>200 visible photons per 8 keV X-ray) and an emission spectrum matched to the typical CCD's spectral sensitivity peak in the red. Although there is a fairly prompt emission of most of the light (<1 ms to 10%), there is a long-term persistence which decays according to a power law: bright spots glow for seconds after an exposure has ended. This severely limits the dynamic range during fast framing, such as might be encountered in a synchrotron environment.

Other dopants for Gd2O2S, such as Tb and Pr, have much shorter persistence and are better suited to higher frame rates. These phosphors yield somewhat lower CCD signals because they emit fewer visible photons per X-ray and because their blue–green emission is less well matched to the CCD spectral sensitivity. Interestingly, Gd2O2S:Tb has one of the slowest `prompt' emission (exponential decay) time constants, but is one of the fastest phosphors to decay to 10−4 (<10 ms), resulting in low persistence.

Thicker phosphors are needed at higher X-ray energies (>15 keV) to maintain high stopping power. This generally reduces spatial resolution. However, structured phosphors offer a way to increase thickness while limiting the lateral spread of the light. For example, CsI:Tl can be grown as an array of columnar crystals, resulting in a screen with enhanced resolution (Stevels & Schrama-de Pauw, 1974a[link], b[link]; Moy, 1998[link]). The spectral mismatch of this green-emitting phosphor is offset by the increased signal per X-ray expected at these higher X-ray energies. Recently, a (Zn,Cd)Se phosphor has been described that has excellent characteristics for X-ray energies above 12 keV; the stopping power at lower energies is compromised by absorption edges (Bruker AXS Inc.).

Fibre optics and lenses. Light transmission in image reduction is limited by [\hbox{n.a.}\times M^{2}], where n.a. is the numerical aperture of the optical system and M is the linear magnification factor ([M \lt 1] corresponds to reduction). Lens systems have low values of n.a. when used in reduction, so there is only 2% light transmission for a 3:1 reduction with an f/1.0 lens. With a much higher n.a., fibre optics can typically transmit 13% with the same 3:1 reduction (Coleman, 1985[link]). Such reducing tapers are produced by locally heating, then pulling, a fused bundle of optical fibres. Each fibre within the bundle becomes tapered in this process, thereby reducing the image scale from front to back of the bundle. The bundle structure introduces a characteristic `chicken-wire' pattern into the image which must be removed via intensity calibration (Section 7.2.3[link]). Tapers up to 165 mm in diameter are available.

To obtain good resolution, extramural absorbing fibres (EMAs) must be placed in the fibre optic to absorb light that propagates between fibres. These EMAs are often a more effective absorber in the blue part of the spectrum, with the result that a red-emitting phosphor yields somewhat poorer resolution. Another concern with fibre optics is radioactivity in the glass used to produce the optic, which manifests itself as random flashes (`zingers') in both the phosphor and the CCD (Section 7.2.3[link]).

Although fibre-optic coupling is preferred in many situations, lenses are appropriate for use with image intensification, or in cases of image magnification, such as for microtomography, using high-resolution screens.

Image intensifiers allow large demagnification ratios without undue DQE loss (Moy, 1994[link]). In these vacuum-tube devices, visible light produces photoelectrons in the photocathode of the intensifier, which in turn are accelerated under high potential onto a secondary phosphor screen. Each photoelectron excites many photons in the secondary phosphor, giving light amplification. Image resolution is retained either by magnetic, electrostatic or proximity focusing of the electrons. In the case of microchannel plate intensifiers, photoelectrons are restricted to cascade down hollow fibres. Intensifiers often have problems with stability and linearity, and they degrade resolution. Phosphor afterglow in the secondary phosphors is also a consideration. Intensifiers typically limit the input format to less than 80 mm in diameter, although one design uses a large (230 mm) radiographic intensifier (Moy, 1994[link]; Hammersley et al., 1997[link]). The greatest drawbacks of image intensifiers are cost, availability, susceptibility to magnetic fields and lack of robustness.

CCDs. The low noise and high sensitivity of scientific CCDs have made CCDs the preferred imaging device in X-ray applications. Visible photons are converted to charge carriers in the silicon of the CCD, with a 30–40% quantum efficiency (q.e.) in the red, dropping to 5% in the blue owing to increased absorption by the controlling gate structure (the q.e. can be considerably higher for back-illuminated devices which, however, are expensive, fragile and difficult to obtain). The `read noise' (noise with zero charge in the pixel) is routinely less than 10 electrons for scientific grade chips with pixel read rates of less than 1 MHz. When coupled to a good phosphor screen with a modest (circa 3:1) fibre-optic taper, signal levels in the CCD are typically 10–30 recorded electrons per 10 keV X-ray. This is above the CCD read-noise level, thereby allowing low-dose quantum-limited X-ray imaging.

A physically large CCD chip is usually desired to maximize the detector area, given the limited image reduction ratio. Devices from 2–4 cm on an edge with 1000 × 1000 to 2000 × 2000 pixels are available. These sizes match well to the largest of the fibre-optic tapers and are therefore most often used in detectors. Larger-format CCDs are available, but are harder to obtain and are expensive.

CCDs are usually cooled well below room temperature to minimize thermally generated dark current, an unwanted source of noise. The dark current drops by a factor of 2 for every 5–7 K drop in temperature. At 233 K, there is roughly one electron per pixel per second dark current for normal clocking operation. Since the dark current changes with temperature, the temperature must be well regulated (to within [\pm]0.1 K to subtract the dark charge from an image reproducibly.

Most of the dark current comes from surface defects, not from the bulk silicon. Multiphase pinned (MPP) CCDs use charge implants within the pixel structure to move the charge collection region away from the surface, thereby reducing the dark current by several orders of magnitude. This is usually accompanied by a significant reduction in the maximum charge capacity (`full well') of a pixel. Even so, MPP CCDs are becoming the norm in most X-ray detectors.

Directly exposed CCDs. CCDs can directly image X-rays, although typical CCDs are not very efficient, since the charge collection region in the silicon is very thin (< 10 µm). Thicker depletion regions can be fabricated in high-resistivity silicon wafers, improving X-ray collection efficiency to greater than 30% at 8 keV. Direct conversion of X-rays in silicon produces a large signal with excellent energy resolution. It is therefore possible to use the chip as an X-ray counting detector with the ability to discriminate in energy. To retain X-ray-energy measurement capability, however, there must be less than one X-ray per pixel per frame to avoid signal overlap. This requires a fast readout as well as large amounts of disk storage to handle the large number of files. The X-rays damage the CCD's electronic structure, resulting in a higher dark current within the exposed pixels. Care must be taken to shield non-imaging parts of the CCD, such as the output amplifier, which will adversely affect chip performance at even lower radiation dose.

References

First citation Coleman, C. I. (1985). Imaging characteristics of rigid coherent fiber optic tapers. Adv. Electron. Electron Phys. 64B, 649–661.Google Scholar
First citation Deckman, H. W. & Gruner, S. M. (1986). Format alterations in CCD based electro-optic X-ray detectors. Nucl. Instrum. Methods Phys. Res. A, 246, 527–533.Google Scholar
First citation Hammersley, A. P., Brown, K., Burmeister, W., Claustre, L., Gonzalez, A., McSweeney, S., Mitchell, E., Moy, J.-P., Svensson, S. O. & Thompson, A. W. (1997). Calibration and application of an X-ray image intensifier/charge-coupled device detector for monochromatic macromolecular crystallography. J. Synchrotron Rad. 4, 67–77.Google Scholar
First citation Moy, J.-P. (1994). A 200 mm input field 5–80 keV detector based on an X-ray image intensifier and CCD camera. Nucl. Instrum. Methods Phys. Res. A, 348, 641–644.Google Scholar
First citation Moy, J.-P. (1998). Image quality of scintillator based X-ray electronic imagers. Proc. SPIE, 3336, 187–194.Google Scholar
First citation Naday, I., Ross, S., Kanyo, M., Westbrook, M., Westbrook, E. M., Phillips, W. C., Stanton, M. J. & O'Mara, D. (1995). The gold detector: modular CCD area detector for macromolecular crystallography. Proc. SPIE, 2415, 236–249.Google Scholar
First citation Shepherd, J. A., Gruner, S. M., Tate, M. W. & Tecotzky, M. (1995). A study of persistence in gadolinium oxysulfide X-ray phosphors. Proc. SPIE, 2519, 24–30.Google Scholar
First citation Stevels, A. & Schrama-de Pauw, A. (1974a). Vapour-deposited CsI:Na layers. I. Morphologic and crystallographic properties. Philips Res. Rep. 29, 340–352.Google Scholar
First citation Stevels, A. & Schrama-de Pauw, A. (1974b). Vapour-deposited CsI:Na layers. II. Screens for application in X-ray imaging devices. Philips Res. Rep. 29, 353–362.Google Scholar
First citation Tate, M. W., Eikenberry, E. F. & Gruner, S. M. (1997). Coupling format variations in X-ray detectors based on charge-coupled devices. Rev. Sci. Instrum. 68, 47–54.Google Scholar








































to end of page
to top of page