Section 6.2.1. Reactors

B. P. Schoenborna* and R. Knottb

a Life Sciences Division M888, University of California, Los Alamos National Laboratory, Los Alamos, NM 8745, USA, and bSmall Angle Scattering Facility, Australian Nuclear Science & Technology Organisation, Physics Division, PMB 1 Menai NSW 2234, Australia
Correspondence e-mail:  schoenborn@lanl.gov

6.2.1. Reactors

| top | pdf |

The generation of neutrons by steady-state nuclear reactors is a well established technique (Bacon, 1962[link]; Kostorz, 1979[link]; Pynn, 1984[link]; Carpenter & Yelon, 1986[link]; Windsor, 1986[link]; West, 1989[link]). Reactor sources that play a major role in neutron-beam applications have a maximum unperturbed thermal neutron flux, [\varphi_{\rm th}], within the range [1 \lt \varphi_{\rm th} \lt 20 \times 10^{14}\;\hbox{cm}^{-2} \hbox{ s}^{-1}]. A research reactor is essentially a matrix of fuel, coolant, moderator and reflector in a well defined geometry (Fig. 6.2.1.1[link]). The fuel is uranium, and neutron-induced fission in the isotope 235U produces a number of prompt and delayed neutrons (slightly more than a total of two) and, on average, one of these is required to maintain the steady-state chain reaction (i.e. criticality). The heat generated (~200 MeV per fission event) must be removed, hence the need for an efficient coolant. In practice, the simultaneous requirements of complex thermal hydraulics and nuclear reaction kinetics must be addressed. The neutrons produced in the fission event have a mean energy of ~1 MeV, and a material is required to reduce this energy to ~25 MeV to take advantage of the larger fission cross section of 235U in the `thermal' energy range. Such a moderator is composed of a material rich in light nuclei, so that a large fraction of the neutron energy is transferred per collision.

[Figure 6.2.1.1]

Figure 6.2.1.1 | top | pdf |

Schematic of the High Flux Beam Reactor at the Brookhaven National Laboratory (USA). The central core region (48 cm in diameter and 58 cm high) contains 28 fuel elements in an array surrounded by an extended D2O moderator/reflector region. The diameter of the reactor vessel is approximately 2 m. All but one of the beam tubes are tangentially oriented with respect to the core. The cold neutron source is located in the H9 beam tube.

There is an inherent maximum in neutron flux density imposed by the fission process (the number of excess neutrons produced per fission event), by the reduced density of neutron-generation material required for cooling purposes and by the heat-removal capacity of suitable coolants. Detailed design of reactor systems is essential to obtain the correct balance.

6.2.1.1. Basic reactor physics

| top | pdf |

Reactor physics is the theoretical and experimental study of the neutron distributions in the energy, spatial and time domains (Soodak, 1962[link]; Jakeman, 1966[link]; Akcasu et al., 1971[link]; Glasstone & Sesonske, 1994[link]). The fundamental relation describing neutron kinetics is the Boltzmann transport equation (e.g. Spanier & Gelbard, 1969[link]; Stamm'ler & Abbate, 1983[link]; Weisman, 1983[link]; Lewis & Miller, 1993[link]). In theory, the transport equation describes the life of the neutron from its birth as a high-energy component of the fission process, through the various diffusion and moderation processes, until its ultimate end in (i) the chain reaction, (ii) leakage into beam tubes, or (iii) parasitic absorption (Williams, 1966[link]). In practice, the complex nuclear reactions and the geometrical configurations of the component materials are such that a rigorous theoretical analysis is not always possible, and simplifying approximations are necessary. Nevertheless, well proven algorithms have been developed, and many have been included in computer codes (e.g. Hallsall, 1995[link]).

On examination of the factors that influence the neutron flux distribution, there are three distinct but interdependent functions performed by the coolant, moderator and reflector. The coolant/moderator is of major importance in the fuelled region to sustain optimum conditions for the chain reaction, and the moderator/reflector is important in the regions surrounding the central core (Section 6.2.1.2[link]). It should be noted that reactors for neutron-beam applications must be substantially undermoderated in order to provide a fast neutron flux at the edge of the core, which can be thermalized at the entry to the beam tubes. Most research reactors use H2O or D2O as the coolant/moderator.

6.2.1.2. Moderators for neutron scattering

| top | pdf |

The moderator/reflector serves to modify the energy distribution of fast neutrons leaking from the central core, returning a significant number of thermalized neutrons to the core region to provide for criticality with a smaller inventory of fuel and providing excess neutrons for a range of applications, including neutron-beam applications. The moderator/reflector may be H2O, D2O, Be, graphite or a combination of these. Almost all choices have been used; however, the optimum is not achieved with any choice, and priorities must be set in terms of neutron-beam performance, other source activities and the reactor fuel cycle.

For neutron-beam applications, H2O is the preferred reflector material since the moderation length is large and the absorption cross section low. Consequently, the thermal neutron flux peak is rather broad and occurs relatively distant from the core. While the exact position of the peak depends on the reactor core design, the peak width has a major impact on beam-tube orientation (Section 6.2.1.3[link]). The combination of D2O coolant/moderator–D2O moderator/reflector provides a distinct advantage; however, for a number of technical reasons, the H2O–D2O combination is becoming more common, with the D2O in a closed vessel surrounding the central H2O-cooled core.

6.2.1.2.1. Thermal moderators

| top | pdf |

Neutrons thermalized within a `semi-infinite' moderator/reflector typical of a steady-state reactor source establish an equilibrium Maxwellian energy distribution characterized by the temperature (T) of the moderator (Fig. 6.2.1.2[link]). The wavelength, [\lambda_{m}], at which the above distribution has a maximum is given by [\lambda_{m} = h/(5k_{B} Tm_{n})^{1/2}.] Depending on the width of the moderator and its composition, the Maxwellian distribution merges with the [1/E] slowing-down distribution from the reactor core to give a total distribution at the beam-tube entry.

[Figure 6.2.1.2]

Figure 6.2.1.2 | top | pdf |

Neutron wavelength distributions for a thermal (310 K) and a cold (30 K) neutron moderator in a `typical' dedicated beam reactor. The Maxwellian distribution merges with a [1/E] slowing-down distribution at shorter wavelengths. A wavelength distribution for a monochromatic beam application on a thermal source is illustrated. It should be noted that, depending on the value of the mean wavelength for the monochromatic beam, harmonic contamination may be significant.

Clearly, the neutron wavelength distribution will depend on the local equilibrium conditions. Since steady-state reactors typically operate with moderator/reflector temperatures in the range 308–323 K, the corresponding [\lambda_{m}] is about 1.4 Å (Fig. 6.2.1.2[link]). However, it is possible to alter the neutron distribution by re-thermalizing the neutrons in special moderator regions, which are either cooled significantly below or heated significantly above the average moderator temperature. One such device of prime importance is the cold moderator in the form of a cold source.

6.2.1.2.2. Cold moderators

| top | pdf |

The thermal neutron distribution shown in Fig. 6.2.1.2[link] is not ideal for all experiments, since the flux of 5 Å neutrons is almost two orders of magnitude less than the peak. The solution is to introduce a cold region in the moderator/reflector. This is typically a volume of liquid H2 or D2 at ~10–30 K, and a Maxwellian distribution around this temperature results in a [\lambda_{m}] of ~4 Å. The geometric design of a cold-source vessel has been shown to be very important, with substantial gains in neutron flux obtained by innovative design. A re-entrant geometry approximately 20 cm in diameter filled with liquid D2 at 10 K provides optimum neutron thermalization with superior coupling to neutron guides (Section 6.2.1.3.5[link]) (Ageron, 1989[link]; Lillie & Alsmiller, 1990[link]; Alsmiller & Lillie, 1992[link]).

6.2.1.3. Beamline components

| top | pdf |

Until recently, instrument design has been largely based on experience; however, in many cases, it is now possible to formulate a comprehensive description of the instrument and explore the impact of various parameters on instrument performance using an extensive array of computational methods (Johnson & Stephanou, 1978[link]; Sivia et al., 1990[link]; Hjelm, 1996[link]). In practice, it is the instrument design that provides access to the fundamental scattering processes, as briefly outlined in the following.

If a neutron specified by a wavevector [{\bf k}_{1}] is incident on a sample with a scattering function [S({\bf Q}, \omega)], all neutron scattering can be reduced to the simple form [{\hbox{d}^{2} \sigma \over \hbox{d}\Omega \hbox{ d}E} = AS({\bf Q}, \omega),] where A is a constant containing experimental information, including instrumental resolution effects. The basic quantity to be measured is the partial differential cross section, which gives the fraction of neutrons of incident energy E scattered into an element of solid angle Ω with an energy between [E'] and [E' + \hbox{d}E']. The momentum transfer, Q, is given in Fig. 6.2.1.3(a[link]). The primary aim of a neutron-scattering experiment is to measure [{\bf k}_{2}] to a predetermined precision (Bacon, 1962[link]; Sears, 1989[link]) (Fig. 6.2.1.3b[link]). A generic neutron-scattering instrument used to achieve this aim is illustrated in Fig. 6.2.1.4[link]. The instrument resolution function will be determined by uncertainties in [{\bf k}_{1}] and [{\bf k}_{2}], which are a direct consequence of (i) measures to increase the neutron flux at the sample position (to maximize wavelength spread, beam divergence, monochromator mosaic, for example) and (ii) uncertainties in geometric parameters (flight-path lengths, detector volume etc.).

[Figure 6.2.1.3]

Figure 6.2.1.3 | top | pdf |

(a) Schematic vector diagram for an elastic neutron-scattering event. A neutron, [{\bf k}_{1}], is incident on a sample, S, and a scattered neutron, [{\bf k}_{2}], is observed at an angle 2θ leading to a momentum transfer, Q. (b) Schematic of an elastic neutron-scattering event illustrating the consequences of uncertainty in defining the incident neutron, [{\bf k}_{1}], and determining the scattered neutron, [{\bf k}_{2}]. The volumes ([\delta {\bf k}_{1_{x}}], [\delta {\bf k}_{1_{y}}], [\delta {\bf k}_{1_{z}}]) and ([\delta {\bf k}_{2_{x}}], [\delta {\bf k}_{2_{y}}], [\delta {\bf k}_{2_{z}}]) constitute the instrument resolution function.

[Figure 6.2.1.4]

Figure 6.2.1.4 | top | pdf |

A generic neutron-scattering instrument illustrating the classes of facilities and operators important to instrument design and assessment. Each class should be optimized and integrated into the overall instrument description.

6.2.1.3.1. Collimators and filters

| top | pdf |

In general, neutron-beam applications are flux-limited, and a major advantage will be realized by adopting advanced techniques in flux utilization. A reactor neutron source is large, and stochastic processes dominate the generation, moderation and general transport mechanisms. Because of radiation shielding, background reduction and space requirements for scattering instruments, reactor beam tubes have a minimum length of 3–4 m and cannot exceed a diameter of about 30 cm. The useful flux at the beam-tube exit is thus between 10−5 and 10−6 times the isotropic flux at its entry.

The most important aspect of beam-tube design, other than the size, is the position and orientation of the tube with respect to the reactor core. The widespread utilization of D2O reflectors enables significant gains to be obtained from tangential tubes with maximal thermal neutron flux, and minimal fast neutron and γ fluxes. A similar result is accomplished in a split-core design by orienting the beam tubes toward the unfuelled region of the reactor core (Prask et al., 1993[link]).

There are two major groups of neutron-scattering instruments: those located close to and those located some distance from the neutron source. The first group is located to minimize the impact of the inverse square law on the neutron flux, and the second group is located to reduce background and provide more instrument space. In both cases, the first beamline component is a collimator to extract a neutron beam of divergence α from the reactor environment. The collimator is usually a beam tube of suitable dimensions for fully illuminating the wavelength-selection device. The angular acceptance of a collimator is determined strictly by the line-of-sight geometry between the source and the monochromator. Some geometric focusing may be appropriate, and a Soller collimator may be used to reduce α without reducing the beam dimensions. For technical reasons, the primary collimator is essentially fixed in dimensions and secondary collimators of adjustable dimensions may be required in more accessible regions outside the reactor shielding.

Neutron-beam filters are required for two main reasons: (i) to reduce beam contamination by fast neutron and γ radiation and (ii) to reduce higher- or lower-order harmonics from a monochromatic beam. Numerous single-crystal, polycrystalline and multilayer materials with suitable characteristics for filter applications are available (e.g. Freund & Dolling, 1995[link]).

6.2.1.3.2. Crystal monochromators

| top | pdf |

The equilibrium neutron-wavelength distribution (Fig. 6.2.1.2[link]) is a broad continuous distribution, and in most experiments it is necessary to select a narrow band in order to define [{\bf k}_{1}]. Neutron-wavelength selection can be achieved by Bragg scattering using single crystals to give well defined wavelengths; by polycrystalline material to remove a range of wavelengths; by a mechanical velocity selector; or by time-of-flight methods. The method chosen will depend on experimental requirements for the wavelength, λ, and wavelength spread, [\Delta \lambda/\lambda].

Neutrons incident on a perfect single crystal of given interplanar spacing d will be diffracted to give specific wavelengths at angle 2θ according to the Bragg relation. A neutron beam, [\alpha_{1}], incident on a single crystal of mosaicity β will provide [\Delta \lambda /\lambda = [(\cot \theta \cdot \theta)^{2} + (\Delta d/d)^{2}]^{1/2},] where [(\Delta \theta)^{2} = {\alpha_{1}^{2} \alpha_{2}^{2} + \beta^{2} (\alpha_{1}^{2} + \alpha_{2}^{2}) \over \alpha_{1}^{2} + \alpha_{2}^{2} + 4\beta^{2}}] and [\alpha_{2}] is the divergence of the (unfocused) diffracted beam.

Crystals for neutron monochromators must not only have a suitable d, but also high reflectivity and adequate β. Under these conditions, neutron beams with a [\Delta \lambda/\lambda] of a few per cent are obtained. Typical crystals are Ge, Si, Cu and pyrolytic graphite. In order to increase the neutron flux at the sample, a number of mechanisms have been developed. These include focusing monochromatic crystals, frequently using Si or Ge (Riste, 1970[link]; Mikula et al., 1990[link]; Copley, 1991[link]; Magerl & Wagner, 1994[link]; Popovici & Yelon, 1995[link]), as well as stacked composite wafer monochromators (Vogt et al., 1994[link]; Schefer et al., 1996[link]).

One limitation of the use of crystal monochromators is the absence of suitable materials with large d. Indeed, the longest useable λ diffracted from pyrolytic graphite or Si is ~5–6 Å.

6.2.1.3.3. Multilayer monochromators and supermirrors

| top | pdf |

Multilayers are especially useful for preparing a long-wavelength neutron beam from a cold source and for small-angle scattering experiments in which [\Delta \lambda /\lambda] of about 0.1 is acceptable (Schneider & Schoenborn, 1984[link]). Multilayer monochromators are essentially one-dimensional crystals composed of alternating layers of neutron-different materials (e.g. Ni and Ti) deposited on a substrate of low surface roughness. In order to produce multilayers of excellent performance, uniform layers are required with low interface roughness, low interdiffusion between layers and high scattering contrast. Various modifications (e.g. carbonation, partial hydrogenation) to the pure Ni and Ti bilayers improve the performance significantly by fine tuning the layer uniformity and contrast (Mâaza et al., 1993[link]). The minimum practical d-spacing is ~50 Å and a useful upper limit is ~150 Å. Multilayer monochromators have high neutron reflectivity (>0.95 is achievable), and their angular acceptance and bandwidth can be selected to produce a neutron beam of desired characteristics (Saxena & Schoenborn, 1977[link], 1988[link]; Ebisawa et al., 1979[link]; Sears, 1983[link]; Schoenborn, 1992a[link]).

The supermirror, a development of the multilayer monochromator concept, consists of a precise number of layers with graded d-spacing. Such a device enables the simultaneous satisfaction of the Bragg condition for a range of λ and, hence, the transmission of a broader bandwidth (Saxena & Schoenborn, 1988[link]; Hayter & Mook, 1989[link]; Böni, 1997[link]).

Polarizing multilayers and supermirrors (Schärpf & Anderson, 1994[link]) facilitate valuable experimental opportunities, such as nuclear spin contrast variation (Stuhrmann & Nierhaus, 1996[link]) and polarized neutron reflectometry (Majkrzak, 1991[link]; Krueger et al., 1996[link]). Supermirrors consisting of Co and Ti bilayers display high contrast for neutrons with a magnetic moment parallel to the saturation magnetization and very low contrast for the remainder. With suitable modification of the substrate to absorb the antiparallel neutrons, a polarizing supermirror will produce a polarized neutron beam (polarization >90%) by reflection.

6.2.1.3.4. Velocity selectors

| top | pdf |

The relatively low speed of longer-wavelength neutrons (~600 m s−1 at 6 Å) enables wavelength selection by mechanical means (Lowde, 1960[link]). In general, there are two classes of mechanical velocity selectors (Clark et al., 1966[link]). Rotating a group of short, parallel, curved collimators about an axis perpendicular to the beam direction will produce a pulsed neutron beam with λ and [\Delta \lambda/\lambda] determined by the speed of rotation. This is a Fermi chopper. An alternate method is to translate short, parallel, curved collimators rapidly across the neutron beam, permitting only neutrons with the correct trajectory to be transmitted. This is achieved in the helical velocity selector, where the neutron wavelength is selected by the speed of rotation and [\Delta \lambda/\lambda] can be modified by changing the angle between the neutron beam and the axis of rotation (Komura et al., 1983[link]). The neutron beam is essentially continuous, the resolution function is approximately triangular and the overall neutron transmission efficiency exceeds 75% in modern designs (Wagner et al., 1992[link]).

6.2.1.3.5. Neutron guides

| top | pdf |

In order for a collimator to be effective, its walls must absorb all incident neutrons. The angular acceptance is strictly determined by the line-of-sight geometry. Neutron guides can be used to improve this acceptance dramatically and to transport neutrons with a given angular distribution, almost without intensity loss, to regions distant from the source (Maier-Leibnitz & Springer, 1963[link]). The basic principle of a guide is total internal reflection. This occurs for scattering angles less than the critical angle, [\theta_{c}], given by [\theta_{c} = 2(1 - n)^{1/2},] where n is the (neutron) index of refraction related to the coherent scattering length, b, of the wall material, viz, [n = 1 - (\lambda^{2} \rho b/2\pi),] where ρ is the atom number density (in [\hbox{cm}^{-3}]). Among common materials, Ni with [b = 1.03 \times 10^{-12}\;\hbox{cm}], in combination with suitable physico-chemical properties, provides the best option, with a critical angle [\theta_{c} = 0.1 \lambda] (in Å). The dependence of [\theta_{c}] on λ implies that guides are more effective for long-wavelength neutrons. With the introduction of supermirror guides with up to four times the [\theta_{c}] of bulk Ni, both thermal and cold neutron beams are being transported and focused with high efficiency (Böni, 1997[link]).

While a straight guide transports long wavelengths efficiently, it continues to transport all neutrons within the critical angle, including non-thermal neutrons emitted within the solid angle of the guide. This situation may be modified significantly by introducing a curvature to the guide. Since a curved neutron guide provides a form of spectral tailoring (cutoff or bandpass filters), simulation is a distinct advantage in exploring the impact of guide geometry on neutron-beam quality (van Well et al., 1991[link]; Copley & Mildner, 1992[link]; Mildner & Hammouda, 1992[link]).

6.2.1.4. Detectors

| top | pdf |

The detection of thermal neutrons is a nuclear event involving one of only a few nuclei with a sufficiently large absorption cross section (3He, 10B, 6Li, Gd and 235U). The secondary products (fragments, charged particles or photons) from the primary nuclear event are used to determine the location. Depending on the geometry of the instrument, either a spatially integrating or a position-sensitive detector is required (Convert & Forsyth, 1983[link]; Crawford, 1992[link]; Rausch et al., 1992[link]). The relatively weak neutron source is driving instrument design towards maximizing the number of neutrons collected per unit time, and, in many cases, this leads to the use of multiwire or position-sensitive detectors. The main performance characteristics for detector systems are position resolution, number of resolution elements, efficiency, parallax, maximum count rate, dynamic range, sensitivity to γ background and long-term stability.

6.2.1.4.1. Multiwire proportional counters

| top | pdf |

The principles of a multiwire proportional counter (MWPC) are well established (Sauli, 1977[link]) and have wide application. For thermal neutron detection (Radeka et al., 1996[link]), the reaction of choice is [^{3}\hbox{He} + \hbox{n} \rightarrow {}^{3}\hbox{H} + \hbox{p} + 764\hbox{ keV}.]

The 191 keV triton and the 573 keV proton are emitted in opposite directions and create a charge cloud whose dimensions are determined primarily by the pressure of a stopping gas. Depending on the work function of the gas mixture, approximately [3\times 10^{4}] electron–ion pairs are created. Low-noise gas amplification of this charge cloud occurs in an intense electric field created in the vicinity of the small diameter (20–30 µm) anode wires (Radeka, 1988[link]). Typical gas gains of ~10–50 lead to a total charge on the anode of ~50–100 fC. The efficiency of the detector is determined by the pressure of 3He, and the spatial resolution and count-rate capability are determined by the detector geometry and readout system. The event decoding is selected from the time difference (Borkowski & Kopp, 1975[link]), charge division (Alberi et al., 1975[link]), centroid-finding filter (Radeka & Boie, 1980[link]), or wire-by-wire techniques (Jacobé et al., 1983[link]; Knott et al., 1997[link]). Present MWPC technology offers opportunities and challenges to design a detector system that is totally integrated into the instrument design and optimizes data collection rate and accuracy (Schoenborn et al., 1985[link], 1986[link]; Schoenborn, 1992b[link]).

A concept related to the MWPC is the micro-strip gas chamber (MSGC). With the MSGC, the general principles of gas detection and amplification apply; however, the anode is deposited on a suitable substrate (Oed, 1988[link], 1995[link]; Vellettaz et al., 1997[link]). The MSGC can potentially improve the performance of the MWPC in some applications, particularly with respect to spatial resolution and count-rate capability.

6.2.1.4.2. Image plates

| top | pdf |

The principles underlying the operation of an image plate (IP) are presented in detail in Chapter 7.2[link] . Briefly, the important difference between an IP for X-ray and neutron detection is the presence of a converter (either Gd2O3 or 6Li). The role of the converter is to capture an incoming neutron and create an event within the IP that mimics the detection of an X-ray photon. For example, neutron capture in Gd produces conversion electrons that exit the Gd2O3 grains, enter neighbouring photostimulated luminescence (PSL) material and create colour centres to form a latent image (Niimura et al., 1994[link]; Takahashi et al., 1996[link]). A neutron IP may have a virtually unlimited area and a shape limited only by the requirement to locate the detection event in a suitable coordinate system. With a neutron-detection efficiency of up to 80% at ~1–2 Å, a dynamic quantum efficiency of ~25–30% can be obtained. The dynamic range is intrinsically 1:105. The spatial resolution is primarily limited by scattering processes of the readout laser beam, and measured line spread functions are typically 150–200 µm. The γ sensitivity is high and may restrict the application to instruments with low ambient γ background.

Neutron IPs are integrating devices well suited to data-acquisition techniques with long accumulation times, such as Laue diffraction (Niimura et al., 1997[link]) and small-angle scattering. On-line readout is a distinct advantage (Cipriani et al., 1997[link]).

6.2.1.5. Instrument resolution functions

| top | pdf |

For accurate data collection, the instrument smearing contribution to the data must be known with some certainty, particularly when data are collected over an extended range with multiple instrument settings. A balance must be struck between instrument smearing and neutron flux at the sample position; however, careful instrument design can produce: (i) a good signal-to-background ratio, thereby partially offsetting the flux limitation, and (ii) facilities and procedures for determining the instrument resolution function (Johnson, 1986[link]).

As an example, instrumental resolution effects in the small-angle neutron scattering (SANS) technique have been investigated in some detail. A `typical' SANS instrument is located on a cold neutron source with an extended (and often variable) collimation system. The sample is as large as possible and the detector is large with low spatial resolution. The instrument is best described by pin-hole geometry. Three major contributions to the smearing of an ideal curve are: (i) the finite λ, (ii) [\Delta \lambda/\lambda] of the beam and (iii) the finite resolution of the detector. Indirect Fourier transform, Monte Carlo and analytical methods have been developed to analyse experimental data and predict the performance of a given combination of resolution-dependent elements (e.g. Wignall et al., 1988[link]; Pedersen et al., 1990[link]; Harris et al., 1995[link]).

References

First citation Ageron, P. (1989). Cold neutron sources at ILL. Nucl. Instrum. Methods A, 284, 197–199.Google Scholar
First citation Akcasu, A. Z., Lellouche, G. S. & Shotkin, L. M. (1971). Mathematical methods in nuclear reactor dynamics. New York: Academic Press.Google Scholar
First citation Alberi, J., Fischer, J., Radeka, V., Rogers, L. C. & Schoenborn, B. P. (1975). A two-dimensional position-sensitive detector for thermal neutrons. Nucl. Instrum. Methods, 127, 507–523.Google Scholar
First citation Alsmiller, R. G. & Lillie, R. A. (1992). Design calculations for the ANS cold source. Part II. Heating rates. Nucl. Instrum. Methods A, 321, 265–270.Google Scholar
First citation Bacon, G. E. (1962). Neutron diffraction. Oxford University Press.Google Scholar
First citation Borkowski, C. J. & Kopp, M. K. (1975). Design and properties of position-sensitive proportional counters using resistance-capacitance position encoding. Rev. Sci. Instrum. 46, 951–962.Google Scholar
First citation Böni, P. (1997). Supermirror-based beam devices. Physica B, 234–236, 1038–1043.Google Scholar
First citation Carpenter, J. M. & Yelon, W. B. (1986). Neutron sources. In Methods of experimental physics, Vol. 23A. New York: Academic Press.Google Scholar
First citation Cipriani, F., Castagna, J.-C., Caustre, L., Wilkinson, C. & Lehmann, M. S. (1997). Large area neutron and X-ray image-plate detectors for macromolecular biology. Nucl. Instrum. Methods A, 392, 471–474.Google Scholar
First citation Clark, C. D., Mitchell, E. W. J., Palmer, D. W. & Wilson, I. H. (1966). The design of a velocity selector for long wavelength neutrons. J. Sci. Instrum. 43, 1–5.Google Scholar
First citation Convert, P. & Forsyth, J. B. (1983). Editors. Position-sensitive detection of thermal neutrons. London: Academic Press.Google Scholar
First citation Copley, J. R. D. (1991). Acceptance diagram analysis of the performance of vertically curved neutron monochromators. Nucl. Instrum. Methods, 301, 191–201.Google Scholar
First citation Copley, J. R. D. & Mildner, D. F. R. (1992). Simulation and analysis of the transmission properties of curved–straight neutron guide systems. Nucl. Sci. Eng. 110, 1–9.Google Scholar
First citation Crawford, R. K. (1992). Position-sensitive detection of slow neutrons – survey of fundamental principles. SPIE, 1737, 210–223.Google Scholar
First citation Ebisawa, T., Achiwa, N., Yamada, S., Akiyoshi, T. & Okamoto, S. (1979). Neutron reflectivities of Ni–Mn and Ni–Ti multilayers for monochromators and supermirrors. J. Nucl. Sci. Technol. 16, 647–659.Google Scholar
First citation Freund, A. K. & Dolling, G. (1995). Devices for neutron beam definition. In International tables for crystallography, Vol. C. Mathematical, physical and chemical tables, edited by A. J. C. Wilson, pp. 375–382. Dordrecht: Kluwer Academic Publishers.Google Scholar
First citation Glasstone, S. & Sesonske, A. (1994). Nuclear reactor engineering. New York: Chapman and Hall.Google Scholar
First citation Hallsall, M. J. (1995). WIMS – a general purpose code for reactor core analysis. AEA Technology, Vienna.Google Scholar
First citation Harris, P., Lebech, B. & Pedersen, J. S. (1995). The three-dimensional resolution function for small-angle scattering and Laue geometries. J. Appl. Cryst. 28, 209–222.Google Scholar
First citation Hayter, J. B. & Mook, H. A. (1989). Discrete thin-film multilayer design for X-ray and neutron supermirrors. J. Appl. Cryst. 22, 35–41.Google Scholar
First citation Hjelm, R. (1996). Editor. Proceedings of the workshop on methods for neutron scattering instrumentation design. Lawrence Berkeley National Laboratory, USA.Google Scholar
First citation Jacobé, J., Feltin, D., Rambaud, A., Ratel, F., Gamon, M. & Pernock, J. B. (1983). High pressure 3He multielectrode detectors for neutron localisation. In Position-sensitive detection of thermal neutrons, edited by P. Convert & J. B. Forsyth, pp. 106–119. London: Academic Press.Google Scholar
First citation Jakeman, D. (1966). Physics of nuclear reactors. London: The English Universities Press.Google Scholar
First citation Johnson, M. W. (1986). Editor. Workshop on neutron scattering data analysis. Rutherford Appleton Laboratory, Chilton, England. Bristol: Institute of Physics.Google Scholar
First citation Johnson, M. W. & Stephanou, C. (1978). MCLIB: a library of Monte Carlo subroutines for neutron scattering problems. Report RL-78–090. Science Research Council, Chilton, England.Google Scholar
First citation Knott, R. B., Smith, G. C., Watt, G. & Boldeman, J. B. (1997). A large 2D PSD for thermal neutron detection. Nucl. Instrum. Methods A, 392, 62–67.Google Scholar
First citation Komura, S., Takeda, T., Fujii, H., Toyoshima, Y., Osamura, K., Mochiki, K. & Hasegawa, K. (1983). The 6-meter neutron small-angle scattering spectrometer at KUR. Jpn. J. Appl. Phys. 22, 351–356.Google Scholar
First citation Kostorz, G. (1979). Neutron scattering. Treatise on materials science and technology, Vol. 15. New York: Academic Press.Google Scholar
First citation Krueger, S., Koenig, B. W., Orts, W. J., Berk, N. F., Majkrzak, C. F. & Gawrisch, K. (1996). Neutron reflectivity studies of single lipid bilayers supported on planar substrates. In Neutrons in biology, edited by B. P. Schoenborn & R. B. Knott, pp. 205–213. New York: Plenum Press.Google Scholar
First citation Lewis, E. E. & Miller, W. F. (1993). Computational methods of neutron transport. Washington: American Nuclear Society Inc.Google Scholar
First citation Lillie, R. A. & Alsmiller, R. G. (1990). Design calculations for the ANS cold neutron source. Nucl. Instrum. Methods A, 295, 147–154.Google Scholar
First citation Lowde, R. D. (1960). The principles of mechanical neutron-velocity selection. J. Nucl. Energy, 11, 69–80.Google Scholar
First citation Magerl, A. & Wagner, V. (1994). Editors. Proceedings of the workshop on focusing Bragg optics. Nucl. Instrum. Methods A, Vol. 338.Google Scholar
First citation Maier-Leibnitz, H. & Springer, T. (1963). The use of neutron optical devices on beam-hole experiments. J. Nucl. Energy, 17, 217–225.Google Scholar
First citation Majkrzak, C. F. (1991). Polarised neutron reflectometry. Physica B, 173, 75–88.Google Scholar
First citation Mikula, P., Krüger, E., Scherm, R. & Wagner, V. (1990). An elastically bent silicon crystal as a monochromator for thermal neutrons. J. Appl. Cryst. 23, 105–110.Google Scholar
First citation Mildner, D. F. R. & Hammouda, B. (1992). The transmission of curved neutron guides with non-perfect reflectivity. J. Appl. Cryst. 25, 39–45.Google Scholar
First citation Mâaza, M., Farnoux, B., Samuel, F., Sella, C., Wehling, F., Bridou, F., Groos, M., Pardo, B. & Foulet, G. (1993). Reduction of the interfacial diffusion in Ni–Ti neutron-optics multilayers by carburation of the Ni–Ti interfaces. J. Appl. Cryst. 26, 574–582.Google Scholar
First citation Niimura, N., Karasawa, Y., Tanaka, I., Miyahara, J., Takahashi, K., Saito, H., Koizumi, S. & Hidaka, M. (1994). An imaging plate neutron detector. Nucl. Instrum. Methods A, 349, 521–525.Google Scholar
First citation Niimura, N., Minezaki, Y., Nonaka, T., Castagna, J.-C., Cipriani, F., Høghøj, P., Lehmann, M. S. & Wilkinson, C. (1997). Neutron Laue diffractometry with an imaging plate provides an effective data collection regime for neutron protein crystallography. Nature Struct. Biol. 4, 909–914.Google Scholar
First citation Oed, A. (1988). Position-sensitive detector with microstrip anode for electron multiplication with gases. Nucl. Instrum. Methods A, 263, 351–359.Google Scholar
First citation Oed, A. (1995). Properties of micro-strip gas chambers (MSGC) and recent developments. Nucl. Instrum. Methods A, 367, 34–40.Google Scholar
First citation Pedersen, J. S., Posselt, D. & Mortensen, K. (1990). Analytical treatment of the resolution function for small-angle scattering. J. Appl. Cryst. 23, 321–333.Google Scholar
First citation Popovici, M. & Yelon, W. B. (1995). Focusing monochromators for neutron diffraction. J. Neutron Res. 3, 1–26.Google Scholar
First citation Prask, H. J., Rowe, J. M., Rush, J. J. & Schroeder, I. G. (1993). The NIST cold neutron research facility. J. Res. NIST, 98, 1–14.Google Scholar
First citation Pynn, R. (1984). Neutron scattering instrumentation at reactor based installations. Rev. Sci. Instrum. 55, 837–848.Google Scholar
First citation Radeka, V. (1988). Low noise techniques in detectors. Annu. Rev. Nucl. Part. Sci. 38, 217–277.Google Scholar
First citation Radeka, V. & Boie, R. A. (1980). Centroid finding method for position-sensitive detectors. Nucl. Instrum. Methods, 178, 543–554.Google Scholar
First citation Radeka, V., Schaknowski, N. A., Smith, G. C. & Yu, B. (1996). High precision thermal neutron detectors. In Neutrons in biology, edited by B. P. Schoenborn & R. B. Knott, pp. 57–67. New York: Plenum Press.Google Scholar
First citation Rausch, C., Bücherl, T., Gähler, R., Seggern, H. & Winnacker, A. (1992). Recent developments in neutron detection. SPIE, 1737, 255–263.Google Scholar
First citation Riste, T. (1970). Singly bent graphite monochromators for neutrons. Nucl. Instrum. Methods. 86, 1–4.Google Scholar
First citation Sauli, F. (1977). Principles of operation of multiwire proportional and drift chambers. Report CERN-77-09. CERN, Geneva, Switzerland.Google Scholar
First citation Saxena, A. M. & Schoenborn, B. P. (1977). Multilayer neutron monochromators. Acta Cryst. A33, 805–813.Google Scholar
First citation Saxena, A. M. & Schoenborn, B. P. (1988). Multilayer monochromators for neutron spectrometers. Mater. Sci. Forum, 27/28, 313–318.Google Scholar
First citation Schefer, J., Medarde, M., Fischer, S., Thut, R., Koch, M., Fischer, P., Staub, U., Horisberger, M., Bottger, G. & Donni, A. (1996). Sputtering method for improving neutron composite germanium monochromators. Nucl. Instrum. Methods A, 372, 229–232.Google Scholar
First citation Schneider, D. K. & Schoenborn, B. P. (1984). A new neutron small-angle diffraction instrument at the Brookhaven High Flux Beam Reactor. In Neutrons in biology, edited by B. P. Schoenborn, pp. 119–141. New York: Plenum Press.Google Scholar
First citation Schoenborn, B. P. (1992a). Multilayer monochromators and super mirrors for neutron protein crystallography using a quasi Laue technique. SPIE, 1738, 192–199.Google Scholar
First citation Schoenborn, B. P. (1992b). Area detectors for neutron protein crystallography. SPIE, 1737, 235–243.Google Scholar
First citation Schoenborn, B. P., Saxena, A. M., Stamm, M., Dimmler, G. & Radeka, V. (1985). A neutron spectrometer with a two-dimensional detector for time resolved studies. Aust. J. Phys. 38, 337–351.Google Scholar
First citation Schoenborn, B. P., Schefer, J. & Schneider, D. (1986). The use of wire chambers in structural biology. Nucl. Instrum. Methods A, 252, 180–187.Google Scholar
First citation Schärpf, O. & Anderson, I. S. (1994). The role of surfaces and interfaces in the behaviour of non-polarizing and polarizing supermirrors. Physica B, 198, 203–212.Google Scholar
First citation Sears, V. F. (1983). Theory of multilayer neutron monochromators. Acta Cryst. A39, 601–608.Google Scholar
First citation Sears, V. F. (1989). Neutron optics: an introduction to the theory of neutron optical phenomena and their applications. Oxford series on neutron scattering in condensed matter. New York: Oxford University Press.Google Scholar
First citation Sivia, D. S., Silver, R. N. & Pynn, R. (1990). The Bayesian approach to optimal instrument design. In Neutron scattering data analysis, edited by M. W. Johnson, Institute of Physics Conference Series, Vol. 107, pp. 45–55.Google Scholar
First citation Soodak, H. (1962). Editor. Reactor handbook. New York: Wiley.Google Scholar
First citation Spanier, J. & Gelbard, E. M. (1969). Monte Carlo principles and neutron transport problems. London: Addison-Wesley.Google Scholar
First citation Stamm'ler, R. J. J. & Abbate, M. J. (1983). Methods of steady-state reactor physics in nuclear design. London: Academic Press.Google Scholar
First citation Stuhrmann, H. B. & Nierhaus, K. H. (1996). The determination of the in situ structure by nuclear spin contrast variation. In Neutrons in biology, edited by B. P. Schoenborn & R. B. Knott, pp. 397–413. New York: Plenum Press.Google Scholar
First citation Takahashi, K., Tazaki, S., Miyahara, J., Karasawa, Y. & Niimura, N. (1996). Imaging performance of imaging plate neutron detectors. Nucl. Instrum. Methods A, 377, 119–122.Google Scholar
First citation Vellettaz, N., Assaf, J. E. & Oed, A. (1997). Two dimensional gaseous microstrip detector for thermal neutrons. Nucl. Instrum. Methods A, 392, 73–79.Google Scholar
First citation Vogt, T., Passell, L., Cheung, S. & Axe, J. D. (1994). Using wafer stacks as neutron monochromators. Nucl. Instrum. Methods A, 338, 71–77.Google Scholar
First citation Wagner, V., Friedrich, H. & Wille, P. (1992). Performance of a high-tech neutron velocity selector. Physica B, 180–181, 938–940.Google Scholar
First citation Weisman, J. (1983). Editor. Elements of nuclear reactor design. Amsterdam: Elsevier Scientific Publishing Company.Google Scholar
First citation Well, A. A. van, de Haan, V. O. & Mildner, D. F. R. (1991). The average number of reflections in a curved neutron guide. Nucl. Instrum. Methods A, 309, 284–286.Google Scholar
First citation West, C. D. (1989). The US advanced neutron source. ICANS X, Los Alamos USA, pp. 643–654.Google Scholar
First citation Wignall, G. D., Christen, D. K. & Ramakrishnan, V. (1988). Instrumental resolution effects in small-angle neutron scattering. J. Appl. Cryst. 21, 438–451.Google Scholar
First citation Williams, M. M. R. (1966). The slowing down and thermalization of neutrons. Amsterdam: North Holland.Google Scholar
First citation Windsor, C. G. (1986). Experimental techniques. In Methods of experimental physics, Vol. 23A. New York, London: Academic Press.Google Scholar








































to end of page
to top of page