International
Tables for
Crystallography
Volume I
X-ray absorption spectroscopy and related techniques
Edited by C. T. Chantler, F. Boscherini and B. Bunker

International Tables for Crystallography (2024). Vol. I. ch. 3.44, pp. 564-566
https://doi.org/10.1107/S1574870722005468

Chapter 3.44. Self-absorption corrections

Frank Bridgesa* and Corwin H. Boothb

aDepartment of Physics, University of California, Santa Cruz, Santa Cruz, CA 95064, USA, and bLawrence Berkeley National Laboratory, Berkeley, CA 94720, USA
Correspondence e-mail:  [email protected]

Currently used self-absorption corrections for extended X-ray absorption fine structure and X-ray absorption near-edge structure fluorescence data are briefly reviewed.

Keywords: self-absorption corrections.

1. Introduction

When all of the atoms of interest in a dilute or very thin sample receive the same incident flux of X-rays, the number of fluorescence photons is proportional to the photoelectric part of the X-ray absorption coefficient μ; see equation (3) in Bridges (2024link to reference). However, many samples are not in this limit: the concentration of the atoms of interest may not be low and the sample may be thick. In this case, the penetration of the X-rays into the sample depends on the total absorption coefficient, and if μ varies rapidly with energy E, as is the case for X-ray absorption near-edge structure (XANES) and extended X-ray absorption fine structure (EXAFS), the mean absorption depth also changes. Then, if μ decreases slightly at some energy, the X-rays penetrate farther into the sample and excite more atoms, and thus the number of fluorescence photons does not decrease as much as expected. This `self-absorption' effect reduces the EXAFS amplitude and can significantly distort the XANES.

Self-absorption has been addressed by several authors over the last few decades (Hayes & Boyce, 1982link to reference; Goulon et al., 1982link to reference; Tan et al., 1989link to reference; Tröger et al., 1992link to reference; Brewe et al., 1994link to reference; Pfalzer et al., 1999link to reference; Booth & Bridges, 2005link to reference; Li et al., 2014link to reference). Hayes & Boyce (1982link to reference) and also Goulon et al. (1982link to reference) briefly discussed errors from self-absorption effects but did not give an explicit correction function. Tan et al. (1989link to reference) and Brewe et al. (1994link to reference) also addressed self-absorption, but took a different approach; they estimated corrections to the amplitude (or coordination number) and the Debye–Waller factor σ2 when self-absorption was important. Tröger et al. (1992link to reference) applied a similar approach to that of Goulon et al. (1982link to reference) for the soft X-ray regime (O K edge) and developed an average correction function to extract the actual EXAFS function χ(k) from the experimental function; however, the correction was for thick samples and did not include changes in μ from the EXAFS oscillations. Booth & Bridges (2005link to reference) extended this calculation to include variable sample thicknesses and also included the effects of the EXAFS oscillations; the latter are important when the EXAFS oscillations are large, such as in ordered materials at low temperatures, for example copper foil. Other approaches have been developed for specialized cases: Li et al. (2014link to reference) considered self-absorption effects in multilayer systems for which refraction and multiple reflections are included, Brewe et al. (1994link to reference) considered a glancing emergent angle geometry and Pfalzer et al. (1999link to reference) considered the case of a large solid-angle detector.

Below, the self-absorption correction function is developed following the approach of Booth & Bridges (2005link to reference); the geometry and some parameters are given in Fig. 1link to figure. Defining y = x/sinφ and z = x/sinθ, and integrating over x, the measured fluorescence intensity If (see Bridges, 2024link to reference) is given byMathematical equationwhere the subscript e refers to the edge of interest, μf is the total absorption coefficient at the fluorescence energy, g = sinφ/sinθ, I0 is the incident intensity and μT = Mathematical symbol, where μb is the slowly changing background absorption from other edges/atoms. Mathematical symbol refers to the average absorption above the edge of interest, which is usually obtained from a spline fit through the EXAFS oscillations. Here, it is assumed that the effects of the finite solid angle of the detector are small enough to be neglected (Booth & Bridges, 2005link to reference); however, see Brewe et al. (1994link to reference) and Pfalzer et al. (1999link to reference) for special cases. The EXAFS function is given by Mathematical symbol and the experimental value by Mathematical symbol. Using equation (1)link to equation plus some algebra (Booth & Bridges, 2005link to reference), χexp can be written in terms of χ asMathematical equationwhere α = Mathematical symbol. At this point the calculation is exact within the assumption of a uniform thickness (plus a uniform distribution of the atom of interest) and a small detector solid angle. The difficulty with it is that χexp is given as a function of χ and cannot be analytically inverted to give χ as a function of χexp for finite sample thicknesses (i.e. αd/sinφ < 3) because of the term Mathematical symbol/sinφ in the exponent. Note that Mathematical symbol and χ is typically less than 0.1. Thus, in most cases Mathematical equationand the exponential term in χ [exp(Mathematical symbol/sinφ)] can be approximated as 1 − (Mathematical symbol/sinφ). Equation (2)link to equation then becomes a quadratic equation in χ, with a solutionMathematical equationwhere Mathematical equation

[Figure 1]

Figure 1

The geometry for calculating the fluorescence output into a small solid angle in EXAFS and XANES experiments. Often both φ and θ are 45°. The point for absorption is a distance x below the surface; y and z are distances along the incoming and outgoing photon paths. Reproduced from Booth & Bridges (2005link to reference). Copyright IOP Publishing. Reproduced with permission. All rights reserved.

The correction factor χ/χexp now oscillates in k-space as a result of the EXAFS oscillations (Booth & Bridges, 2005link to reference). An example of the improvement achieved using the above approximation is shown in Fig. 3 of Booth & Bridges (2005link to reference) for a 4.6 µm foil. Note that equation (2)link to equation can be inverted numerically and hence the approximation (equation 3link to equation) can be checked. This correction has been incorporated into several EXAFS packages such as RSXAP (REDUCE and SABCOR; Booth, 2010link to reference; Booth & Bridges, 2024link to reference) and IFEFFIT (Newville, 2004link to reference; Newville & Ravel, 2024link to reference). For RSXAP, equation (2)link to equation is now inverted numerically.

2. Self-absorption corrections to XANES

The distortion of the XANES from self-absorption is described by the same starting equation as used for EXAFS (equation 1link to equation) after background fluorescence contributions have been subtracted. Also, all treatments consider only the infinitely thick limit, for which the exponential term goes to zero. However, remember that the total absorption coefficient μT is a sum of μe and μb, where μb is the background contribution from other absorption processes. Following Haskel (1999link to reference), an energy EN needs to be chosen above the edge for normalization: it is assumed that this energy is well above the edge and that variations in μe are small (if not, a point should be chosen on Mathematical symbol, for example at a `zero-crossing' in the EXAFS oscillations). The measured absorption is given by If(E)/I0(E), the value of this quantity at the normalization energy is If(EN)/I0(EN) and the normalized absorption is given by μe(E)/μe(EN). Defining the normalized measured absorption Xe as [If(E)/I0(E)]/[If(EN)/I0(EN)], Xe is given by Mathematical equationwhere B = μf/μe(EN), δ = μb(E)/μe(EN) and δ′ = μb(EN)/μe(EN). Solving for μe(E)/μe(EN) and simplifying, one obtains (Haskel, 1999link to reference)Mathematical equationHere, it is assumed that ɛe(E)/ɛe(EN) ≃ 1, and δδ′ for a small energy range around the edge. Note that equation (6)link to equation is normalized (Xe ≃ 1) above the edge energy, where self-absorption effects are large. Consequently, the correction factor in equation (6)link to equation is ∼1.0 at EN and decreases as energy decreases towards the bottom of the edge; i.e. the experimental Xe is suppressed the most for energies near the bottom of the edge, where Xe ≃ 0. At this point on the edge, the extra correction factor is 1/[1 + 1/(Bg + δ)]; as Xe increases, the correction factor approaches 1.0. In contrast, at a white line, particularly for an LIII edge, the correction factor must be greater than 1 because Xe > 1 and the correction factor is 1/[1 − (Xe − 1)/(Bg + δ)]. As long as the correction factor in equation (6)link to equation is less than 3–4 this is typically a good approximation, but since it depends on the quality of the data and the choice of EN it should be checked.

References

First citationBooth, C. H. (2010). RSXAP EXAFS Analysis Package. https://lise.lbl.gov/RSXAP/ .Google Scholar
First citationBooth, C. H. & Bridges, F. (2005). Phys. Scr. 2005, 202.Google Scholar
First citationBooth, C. H. & Bridges, F. (2024). Int. Tables Crystallogr. I, ch. 6.19, 822–826 .Google Scholar
First citationBrewe, D. L., Pease, D. M. & Budnick, J. I. (1994). Phys. Rev. B, 50, 9025–9030.Google Scholar
First citationBridges, F. (2024). Int. Tables Crystallogr. I, ch. 3.41, 546–548 .Google Scholar
First citationGoulon, J., Goulon-Ginet, C., Cortes, R. & Dubois, J. M. (1982). J. Phys. Fr. 43, 539–548.Google Scholar
First citationHaskel, D. (1999). FLUO: Correcting XANES for Self-Absorption in Fluorescence Data. https://www3.aps.anl.gov/haskel/FLUO/Fluo-manual.pdf .Google Scholar
First citationHayes, T. M. & Boyce, J. B. (1982). Solid State Physics, Vol. 37, edited by H. Ehrenreich, F. Seitz & D. Turnbull, p. 173. New York: Academic Press.Google Scholar
First citationLi, W.-B., Yang, X.-Y., Zhu, J.-T., Tu, Y.-C., Mu, B.-Z., Yu, H.-S., Wei, X.-J., Huang, Y.-Y. & Wang, Z.-S. (2014). J. Synchrotron Rad. 21, 561–567.Google Scholar
First citationNewville, M. (2004). IFEFFIT. https://cars9.uchicago.edu/ifeffit/ .Google Scholar
First citationNewville, M. & Ravel, B. (2024). Int. Tables Crystallogr. I, ch. 6.13, 791–795 .Google Scholar
First citationPfalzer, P., Urbach, J.-P., Klemm, M., Horn, S., denBoer, M. L., Frenkel, A. L. & Kirkland, J. P. (1999). Phys. Rev. B, 60, 9335–9339.Google Scholar
First citationTan, Z., Budnick, J. I. & Heald, S. M. (1989). Rev. Sci. Instrum. 60, 1021–1025.Google Scholar
First citationTröger, L., Arvanitis, D., Baberschke, K., Michaelis, H., Grimm, U. & Zschech, E. (1992). Phys. Rev. B, 46, 3283–3289.Google Scholar








































to end of page
to top of page